Complications of Endoscopic Pressure Study Integrated System: Review of 1205 Cases in 10 Years’ Experience

Miyuki Iwasaki, Ippei Tanaka iD, Haruhiro Inoue, Masachika Saino, Kei Ushikubo, Kazuki Yamamoto iD, Yohei Nishikawa, Kaori Owada, Satoshi Abiko iD, Mayo Tanabe

DEN Open · 2026-04 · 6 (1)

DOI: 10.1002/deo2.70173 ↗

Kurz gefasst

Fachartikel in der Zeitschrift DEN Open, online erschienen am 16.07.2025. Verfasst von Miyuki Iwasaki und 9 weiteren Autor:innen. Beteiligte Einrichtung: Digestive Diseases Center Showa University Koto Toyosu Hospital Tokyo Japan. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims The Endoscopic Pressure Study Integrated System (EPSIS) is a novel endoscopic tool to assess lower esophageal sphincter (LES) function by monitoring intragastric pressure (IGP) through insufflation. While previous studies have confirmed its diagnostic utility for LES dysfunction‐related disorders such as gastroesophageal reflux disease and achalasia, there have been no reports evaluating its safety. Thus, this study aimed to assess its complication rates and detailed characteristics.

Methods

We retrospectively reviewed 1205 consecutive EPSIS cases performed between December 2015 and September 2024 at our institution. Cases with adverse events (AEs) were identified, and clinical characteristics, waveform patterns, IGP parameters, and need for additional treatment were analyzed.

Results

AEs occurred in 35 cases (2.9%), including 32 cases (2.7%) of micro‐mucosal bleedings (MMBs) and three cases (0.25%) of superficial mucosal tears (SMTs). All cases were minor (Clavien‐Dindo grade I), without the need for any additional intervention. All 35 cases exhibited an uphill waveform. SMTs were associated with pressure spikes due to vomiting or belching. No serious AEs, including perforations or active bleeding, were observed.

Conclusions

EPSIS is a safe and minimally invasive procedure, with low complication rates comparable to routine upper endoscopy. To maintain its safety, EPSIS should be performed under sedation with CO₂ insufflation. In addition, insufflation during EPSIS should be controlled so that the IGP is maintained less than 25 mmHg. However, this threshold may vary depending on the patient's underlying conditions or clinical background; therefore, further research is warranted. Clinical Trial Registration N/A.

Schlagwörter

Stichwörter der Autor:innen

complicationendoscopic pressure study integrated systemintragastric pressuremicro‐mucosal bleedingssuperficial mucosal tears

Autor:innen und Institutionen

  • Digestive Diseases Center Showa University Koto Toyosu Hospital Tokyo Japan.

ORCID-Profile: Ippei Tanaka, Kazuki Yamamoto, Satoshi Abiko

Zitieren

APA

Iwasaki, M., Tanaka, I., Inoue, H., Saino, M., Ushikubo, K., Yamamoto, K., Nishikawa, Y., Owada, K., Abiko, S., & Tanabe, M. (2026). Complications of Endoscopic Pressure Study Integrated System: Review of 1205 Cases in 10 Years’ Experience. DEN Open, 6(1). https://doi.org/10.1002/deo2.70173

Vancouver

Iwasaki M, Tanaka I, Inoue H, Saino M, Ushikubo K, Yamamoto K, et al. Complications of Endoscopic Pressure Study Integrated System: Review of 1205 Cases in 10 Years’ Experience. DEN Open. 2026;6(1). doi:10.1002/deo2.70173

BibTeX
@article{Iwasaki2026,
  author = {Iwasaki, Miyuki and Tanaka, Ippei and Inoue, Haruhiro and Saino, Masachika and Ushikubo, Kei and Yamamoto, Kazuki and Nishikawa, Yohei and Owada, Kaori and Abiko, Satoshi and Tanabe, Mayo},
  title = {Complications of Endoscopic Pressure Study Integrated System: Review of 1205 Cases in 10 Years’ Experience},
  journal = {DEN Open},
  year = {2026},
  volume = {6},
  number = {1},
  doi = {10.1002/deo2.70173}
}
Kostenlos starten →