Whole blood no‐lyse no‐wash micromethod for the quantitative measurement of monocyte HLA‐DR
Cytometry Part B: Clinical Cytometry · 2024
DOI: 10.1002/cyto.b.22142 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cytometry Part B: Clinical Cytometry, 2024. Verfasst von Jordi Miatello und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Integrative Biology of the Cell, CNRS, CEA, Paris-Saclay University, Gif-sur-Yvette, France. Hauptthemen laut MEDLINE: Monocytes, Critical Illness. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Monocyte (m)HLA‐DR expression appears to be a potent marker of immunosuppression in critically ill patients. The persistence of low mHLA‐DR expression is associated with an increased risk of nosocomial infections and mortality. To adapt this measurement to pediatric requirements and provide extensive 24/7 access, we have developed a whole blood no‐lyse no‐wash micromethod (MM) and compared it with the standardized method (SM). Methods mHLA‐DR was quantified by flow cytometry using Quantibrite™ Anti‐HLA‐DR PE/Monocyte PerCP‐Cy™5.5 with either the SM performed in a diagnostic hematology laboratory using manufacturer protocol, or a whole blood no‐lyse no‐wash MM using an Attune flow cytometer located in the pediatric ICU. Median fluorescence intensity was measured in both techniques and converted to antibodies per cell (AB/C) calibrated with BD Quantibrite™ PE beads. Blood and Quantibrite™ reagent volume used with the MM was reduced by 5‐fold compared to SM. In addition to Quantibrite™ Anti‐Human HLA‐DR PE/Monocyte PerCP‐Cy™5.5, MM required anti‐CD45 and anti‐CD19 labeling.
Results
We determined the expression of mHLA‐DR in 34 patients, 20 adults, and 14 children admitted to ICU. Correlation between MM and SM was excellent (Pearson's correlation: y = 0.8192 x + 678.7, r = 0.9270, p < 0.0001). The estimated bias was 2467 ± 1.96 × 3307 AB/C; CI 95% [−4016; +8949].
Conclusions
The no‐lyse no‐wash whole blood microvolume method for measuring mHLA‐DR expression allows for simplified sample preparation without compromising accuracy of the data. This method may simplify immune monitoring of critically ill patient by the deployment of a point of care method.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Integrative Biology of the Cell, CNRS, CEA, Paris-Saclay University, Gif-sur-Yvette, France.
- Paediatric Intensive Care and Neonatal Medicine, AP-HP, Paris-Saclay University, Bicêtre Hospital, Le Kremlin-Bicêtre, France.
- FHU Sepsis, AP-HP, Paris-Saclay University, INSERM, Le Kremlin-Bicêtre, France.
- INSERM UMR1141 Neurodiderot, Université Paris Cité, Paris, France.
- Saint-Louis Lariboisière Hospital, AP-HP, Denis Diderot University, Paris, France.
- Denis Diderot University, Sorbonne Paris Cité, Paris, France.
ORCID-Profile: Jordi Miatello
Zitieren
APA
Miatello, J., Faivre, V., Marais, C., Raineau, M., Payen, D., & Tissieres, P. (2024). Whole blood no‐lyse no‐wash micromethod for the quantitative measurement of monocyte HLA‐DR. Cytometry Part B: Clinical Cytometry. https://doi.org/10.1002/cyto.b.22142
Vancouver
Miatello J, Faivre V, Marais C, Raineau M, Payen D, Tissieres P. Whole blood no‐lyse no‐wash micromethod for the quantitative measurement of monocyte HLA‐DR. Cytometry Part B: Clinical Cytometry. 2024. doi:10.1002/cyto.b.22142
BibTeX
@article{Miatello2024,
author = {Miatello, Jordi and Faivre, Valérie and Marais, Clémence and Raineau, Mégane and Payen, Didier and Tissieres, Pierre},
title = {Whole blood no‐lyse no‐wash micromethod for the quantitative measurement of monocyte HLA‐DR},
journal = {Cytometry Part B: Clinical Cytometry},
year = {2024},
doi = {10.1002/cyto.b.22142}
}
EUVIMED