Reducing Patient Radiation Skin Dose in Percutaneous Coronary Intervention Using a Dedicated Protocol: A Single‐Center Experience
Catheterization and Cardiovascular Interventions · 09.11.2026
DOI: 10.1002/ccd.71003 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 09.11.2026. Verfasst von Muhammad Ghallab und 12 weiteren Autor:innen. Beteiligte Einrichtung: Creighton University School of Medicine, Omaha, Nebraska, USA. Die Arbeit nennt 29 Literaturangaben.
Abstract (Originalsprache)
Background
Percutaneous coronary interventions (PCI) expose patients and operators to ionizing radiation, posing risks of deterministic effects such as skin injury and stochastic effects including malignancy. Although real‐time dosimetry and dose alerts have been associated with radiation‐dose reductions in prior reports, the level of patient radiation reduction associated with implementing a formal, multifaceted radiation reduction program has not previously been reported in a high‐volume catheterization laboratory.
Objectives
To evaluate the impact of a substantial radiation dose level (SRDL) protocol on patient radiation exposure during PCI.
Methods
This retrospective study analyzed 19,249 PCI procedures at a tertiary center from July 2012 to December 2021. Patients were stratified into a pre‐protocol cohort (Phase 1: 17,332 procedures) and a post‐protocol cohort (Phase 2: 1917 procedures following implementation in November 2019). The SRDL protocol included real‐time radiation monitoring, alert thresholds at 5 Gy air kerma, and structured follow‐up for high‐dose cases. Study outcomes included air kerma ( K a,r ) dose‐area product (DAP), contrast volume, and fluoroscopy time.
Results
The study cohort (mean age 66.3 years; 28.5% female) demonstrated significant reductions in radiation exposure after implementing the protocol. Median air kerma declined from 2.6 to 2.0 Gy (23% reduction; p < 0.0001), and median DAP decreased from 158 to 125 Gy·cm 2 (20.8% reduction; p < 0.0001). Contrast volume was reduced by 15 mL (11.1% reduction; p < 0.0001), while fluoroscopy time remained unchanged (18.1 vs. 18.3 min; p = 0.094). Despite increased procedural complexity in Phase 2 (multivessel PCI: 69.5% vs. 65.2%; p < 0.001), radiation reduction remained significant.
Conclusions
Implementation of an SRDL protocol was associated with significant reductions in patient radiation exposure during PCI without increasing procedure time, supporting its use as a safety intervention in high‐volume catheterization laboratories.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Creighton University School of Medicine, Omaha, Nebraska, USA.
- Icahn School of Medicine at Mount Sinai, New York, New York, USA.
- Chandler Regional Medical Center, Globe, Arizona, USA.
- Gagnon Cardiovascular Institute, Atlantic Health System, Morristown, New Jersey, USA.
- Lenox Hill Hospital, New York, New York, USA.
ORCID-Profile: Muhammad Ghallab, Derek Pineda, Samantha Santori, Bradley Delman, Annapoorna Kini, Samin Sharma, Roxana Mehran, Joseph Sweeny
Zitieren
APA
Ghallab, M., Pineda, D., Bagga, S., Zhang, Z., Giustino, G., Santori, S., Kamen, J., Delman, B., Farhan, S., Kini, A., Sharma, S., Mehran, R., & Sweeny, J. (2026). Reducing Patient Radiation Skin Dose in Percutaneous Coronary Intervention Using a Dedicated Protocol: A Single‐Center Experience. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.71003
Vancouver
Ghallab M, Pineda D, Bagga S, Zhang Z, Giustino G, Santori S, et al. Reducing Patient Radiation Skin Dose in Percutaneous Coronary Intervention Using a Dedicated Protocol: A Single‐Center Experience. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.71003
BibTeX
@article{Ghallab2026,
author = {Ghallab, Muhammad and Pineda, Derek and Bagga, Shiv and Zhang, Zhongjie and Giustino, Gennaro and Santori, Samantha and Kamen, Jacob and Delman, Bradley and Farhan, Serdar and Kini, Annapoorna and Sharma, Samin and Mehran, Roxana and Sweeny, Joseph},
title = {Reducing Patient Radiation Skin Dose in Percutaneous Coronary Intervention Using a Dedicated Protocol: A Single‐Center Experience},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.71003}
}
EUVIMED