Endovascular Bailout Using Covered‐Stent Fenestration After Misdeployment of a Balloon‐Expandable Covered Stent From the Common Iliac Artery Into the Internal Iliac Artery
Catheterization and Cardiovascular Interventions · 02.08.2026
DOI: 10.1002/ccd.70796 ↗Kurz gefasst
Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 02.08.2026. Verfasst von Akinori Satake und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Narita Memorial Hospital, Toyohashi, Aichi, Japan. Hauptthemen laut MEDLINE: Iliac Artery, Angioplasty, Balloon, Stents, Intermittent Claudication. Die Arbeit nennt 3 Literaturangaben.
Abstract (Originalsprache)
Misdeployment of balloon‐expandable covered stents extending from the common iliac artery (CIA) to the internal iliac artery (IIA) is a rare but serious complication. Optimal bailout strategies remain limited. Herein, we report a case of endovascular bailout for inadvertent stent deployment from the CIA to the IIA. A 68‐year‐old man presented with severe claudication of both legs. The bilateral ankle‐brachial index was low, and computed tomography showed chronic total occlusion of the left CIA. During revascularization, a balloon‐expandable covered stent was inadvertently deployed from the CIA into the IIA. To salvage the external iliac artery (EIA), a modified‐tip 0.035‐inch guidewire was used to perforate the stent graft, creating a fenestration. A 0.014‐inch guidewire was advanced through the newly created opening, followed by deployment of a self‐expanding bare‐nitinol stent across the fenestration to reconstruct the EIA flow. The final angiography indicated successful revascularization without complications. Therefore, percutaneous fenestration of a balloon‐expandable covered stent is a potential therapeutic option for the treatment of aortoiliac lesions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Cardiology, Narita Memorial Hospital, Toyohashi, Aichi, Japan.
- Department of Cardiology, Nagoya Heart Center, Nagoya, Aichi, Japan.
- Department of Vascular Surgery, Aichi Medical University, Nagakute, Aichi, Japan.
- Department of Cardiology, Aichi Medical University, Nagakute, Aichi, Japan.
ORCID-Profile: Akinori Satake, Takahiro Tokuda
Zitieren
APA
Satake, A., Tokuda, T., Kodama, A., & Amano, T. (2026). Endovascular Bailout Using Covered‐Stent Fenestration After Misdeployment of a Balloon‐Expandable Covered Stent From the Common Iliac Artery Into the Internal Iliac Artery. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70796
Vancouver
Satake A, Tokuda T, Kodama A, Amano T. Endovascular Bailout Using Covered‐Stent Fenestration After Misdeployment of a Balloon‐Expandable Covered Stent From the Common Iliac Artery Into the Internal Iliac Artery. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70796