Long‐Term Trends in Cardiac Erosion After Transcatheter Closure of Atrial Septal Defects With the Amplatzer Septal Occluder
Catheterization and Cardiovascular Interventions · 2026
DOI: 10.1002/ccd.70722 ↗Kurz gefasst
Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2026. Verfasst von Daniel S Levi und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Mattel Children's Hospital at UCLA, Los Angeles, California, USA. Hauptthemen laut MEDLINE: Heart Septal Defects, Atrial, Cardiac Catheterization, Septal Occluder Device, Prosthesis Failure. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Trends in cardiac erosion after closure of secundum atrial septal defects (ASD) with the Amplatzer Septal Occluder (ASO) have not been assessed since updates were made to the Instructions for Use in 2012 and 2018. This study evaluates the incidence and demographics of ASO erosion over time since its introduction in 1996.
Methods
Data were analyzed for all cardiac erosions reported through March 2025 after ASD closure with an ASO device from 1996 to 2024, as adjudicated and characterized by the AGA Erosion Board or Abbott Medical Affairs. The number of implants annually was estimated as the number of units sold in the calendar year. Overall and time‐dependent erosion rates were calculated as the ratio of reported erosions to the estimated number of implants during each calendar year.
Results
As of March 2025, 198 erosions had been reported and confirmed for ASO devices implanted from 1996 through 2024, which corresponds to an estimated erosion rate of 0.037%. Overall, 31% of erosions occurred within 72 h of implant, 55% within 3 months, and 77% within 1 year, while 13% occurred >5 years after implant. The 1‐month, 1‐year, and 3‐year erosion rates were 0.016%, 0.029%, and 0.031%, respectively. The overall erosion rate decreased significantly from 1996 to 2012 to 2013 to 2024 (0.056% to 0.020%, p < 0.001), as did 3‐day, 1‐month, 1‐year, and 3‐year erosion rates (all p < 0.001). The erosion‐related mortality rate decreased from 0.0040% to 0.0014% from the early era to the later era ( p = 0.061).
Conclusions
Although limitations in the method of estimating implant numbers and ascertaining erosions must be acknowledged, ASO erosion rates and related mortality have decreased significantly since 2012.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Pediatrics, Mattel Children's Hospital at UCLA, Los Angeles, California, USA.
- Departments of Cardiothoracic Surgery and Pediatrics, Stanford University School of Medicine, Palo Alto, California, USA.
Zitieren
APA
Levi, D. S., & McElhinney, D. B. (2026). Long‐Term Trends in Cardiac Erosion After Transcatheter Closure of Atrial Septal Defects With the Amplatzer Septal Occluder. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70722
Vancouver
Levi DS, McElhinney DB. Long‐Term Trends in Cardiac Erosion After Transcatheter Closure of Atrial Septal Defects With the Amplatzer Septal Occluder. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70722
BibTeX
@article{Levi2026,
author = {Levi, Daniel S and McElhinney, Doff B},
title = {Long‐Term Trends in Cardiac Erosion After Transcatheter Closure of Atrial Septal Defects With the Amplatzer Septal Occluder},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70722}
}
EUVIMED