Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 30.12.2025. Verfasst von Feihuang Han und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Guangdong Provincial People's Hospital (Guangdong Academy of Medical Sciences), Southern Medical University, Guangzhou, Chi…. Hauptthemen laut MEDLINE: Device Removal, Heart Arrest, Stents, Angioplasty, Balloon, Coronary. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben. Gefördert von: China International Medical Foundation, Natural Science Foundation of Tibet Autonomous Region.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Balloon catheter fracture is a rare PCI complication that may trigger catastrophic coronary entrapment. This represents the first reported cardiac arrest during percutaneous retrieval of an entrapped, fractured stent balloon in an unprotected left main artery. Case Summary A 72‐year‐old male developed sudden hemodynamic collapse during balloon extraction due to a shaft fracture. Under ongoing CPR with IABP/intubation support, modified balloon anchoring via high‐pressure secondary inflation in 7 F extension catheter enabled en bloc retrieval, with cardiac surgical/ECMO teams preemptively mobilized. Coronary flow was immediately restored, permitting extubation within 1 h and discharge after 5 days without sequelae.

Conclusion

Salvage of such events mandates operator agility for bailout adaptation and on‐site availability of a multidisciplinary rescue team.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Device RemovalHeart ArrestStentsAngioplasty, Balloon, CoronaryEquipment FailureCardiac CathetersCoronary Artery DiseaseHumansMaleAgedTreatment OutcomeCardiopulmonary ResuscitationExtracorporeal Membrane OxygenationHemodynamics

Autor:innen und Institutionen

  • Department of Cardiology, Guangdong Provincial People's Hospital (Guangdong Academy of Medical Sciences), Southern Medical University, Guangzhou, China.
  • Department of Cardiology, National Clinical Research Center for Cardiovascular Diseases, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College,…

ORCID-Profile: Feihuang Han, Zehan Huang, Song Wen

Förderung

  • China International Medical Foundation
  • Natural Science Foundation of Tibet Autonomous Region

Zitieren

APA

Han, F., Ma, D., Wan, Q., Huang, Y., Huang, Z., Wang, F., Wen, S., Zhang, B., & Liu, S. (2026). Bailout Retrieval of an Entrapped, Fractured, and Undeflatable Stent Balloon in the Left Main Coronary Artery: A Salvage Procedure During Cardiac Arrest. Catheterization and Cardiovascular Interventions, 107(4), 1120-1126. https://doi.org/10.1002/ccd.70453

Vancouver

Han F, Ma D, Wan Q, Huang Y, Huang Z, Wang F, et al. Bailout Retrieval of an Entrapped, Fractured, and Undeflatable Stent Balloon in the Left Main Coronary Artery: A Salvage Procedure During Cardiac Arrest. Catheterization and Cardiovascular Interventions. 2026;107(4):1120-1126. doi:10.1002/ccd.70453

BibTeX
@article{Han2026,
  author = {Han, Feihuang and Ma, Dunliang and Wan, Qiheng and Huang, Yuqing and Huang, Zehan and Wang, Feng and Wen, Song and Zhang, Bin and Liu, Sue},
  title = {Bailout Retrieval of an Entrapped, Fractured, and Undeflatable Stent Balloon in the Left Main Coronary Artery: A Salvage Procedure During Cardiac Arrest},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2026},
  volume = {107},
  number = {4},
  pages = {1120--1126},
  doi = {10.1002/ccd.70453}
}
Kostenlos starten →