Comparison of Flow Dynamics After Left Atrial Appendage Occlusion With the Watchman FLX Versus Amulet Devices: Implications for Device‐Related Thrombosis
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31710 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Ahmad Bshennaty und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Michigan Technological University, Houghton, Michigan, USA. Hauptthemen laut MEDLINE: Atrial Appendage, Atrial Fibrillation, Cardiac Catheterization, Thrombosis. Die Arbeit nennt 23 Literaturangaben. Gefördert von: This study was funded by the American Heart Association award #24CDA1269661..
Abstract (Originalsprache)
Background
This study aims to assess how Watchman FLX (Boston Scientific) and Amplatzer Amulet (Abbott) devices for left atrial appendage occlusion (LAAO) impact the flow dynamics in the LA and the risk of device‐related thrombosis (DRT).
Methods
Flow simulations were carried out for 12 AF patients before and after undergoing LAAO. Six patients had Watchman FLX while the other six had Amulet devices, of which three in each group had DRT. The flow parameters investigated included the LA velocities, pressure, time averaged wall shear stress (TAWSS), oscillatory shear index (OSI), and relative residence time (RRT). Low TAWSS and high OSI and RRT are generally associated with flow stasis and thrombosis.
Results
LA flow dynamics differed significantly between the two groups. On average, after LAAO, the Amulet patients exhibited a 25% increase in TAWSS and 24% decrease in RRT while the Watchman FLX patients exhibited, on average, a 24% decrease in TAWSS and a 23% increase in RRT. Both populations experienced an increase in LA pressure (~10% for Amulet patients and 16% for Watchman FLX patients). Increased DRT volume seems to be correlated with low TAWSS and elevated RRT.
Conclusion
This preliminary study suggests that the Amulet device might be associated with improved LA flow dynamics and a potentially lower incidence of DRT compared with the Watchman. However, a larger clinical study is needed to confirm these findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Biomedical Engineering, Michigan Technological University, Houghton, Michigan, USA.
- Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA.
- FEops, Gent, Belgium.
- Department of Cardiology, Aarhus University Hospital, Aarhus, Denmark.
- Department of Cardiology, Copenhagen University Hospital, Copenhagen, Denmark.
- Health Research Institute, Center of Biocomputing and Digital Health and Institute of Computing and cybersystems, Michigan Technological University, Houghton, Michigan, USA.
ORCID-Profile: Alessandra Maria Bavo, Yuheng Jia, Ole De Backer, Hoda Hatoum
Förderung
- This study was funded by the American Heart Association award #24CDA1269661.
Zitieren
APA
Bshennaty, A., Vogl, B. J., Saleh, G., Bavo, A. M., Sularz, A., Ghazal-Aswad, K., Kramer, A. D., Jia, Y., De Beule, M., Nielsen-Kudsk, J. E., De Backer, O., Alkhouli, M., & Hatoum, H. (2025). Comparison of Flow Dynamics After Left Atrial Appendage Occlusion With the Watchman FLX Versus Amulet Devices: Implications for Device‐Related Thrombosis. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31710
Vancouver
Bshennaty A, Vogl BJ, Saleh G, Bavo AM, Sularz A, Ghazal-Aswad K, et al. Comparison of Flow Dynamics After Left Atrial Appendage Occlusion With the Watchman FLX Versus Amulet Devices: Implications for Device‐Related Thrombosis. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31710
BibTeX
@article{Bshennaty2025,
author = {Bshennaty, Ahmad and Vogl, Brennan J and Saleh, Ghasaq and Bavo, Alessandra Maria and Sularz, Agata and Ghazal-Aswad, Karam and Kramer, Anders Dahl and Jia, Yuheng and De Beule, Matthieu and Nielsen-Kudsk, Jens Erik and De Backer, Ole and Alkhouli, Mohamad and Hatoum, Hoda},
title = {Comparison of Flow Dynamics After Left Atrial Appendage Occlusion With the Watchman FLX Versus Amulet Devices: Implications for Device‐Related Thrombosis},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31710}
}
EUVIMED