Percutaneous Management of Vessel Occlusion Caused by Suture‐Based Closure Devices: A Case Series and Benchtop Model
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31598 ↗Kurz gefasst
Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Dan Haberman und 13 weiteren Autor:innen. Beteiligte Einrichtung: Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, USA. Hauptthemen laut MEDLINE: Vascular Closure Devices, Hemostatic Techniques, Catheterization, Peripheral, Suture Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
Large‐bore arterial access (LBAA) closure for percutaneous interventional procedures is often achieved using suture‐based vascular closure devices (VCDs), most commonly Perclose ProGlide (Abbott Vascular, Santa Clara, CA). Although the rate of vascular occlusion due to suture‐based VCDs is rare, there is no widely accepted percutaneous bailout treatment for this device‐related complication.
Aims
This investigation sought to evaluate the percutaneous treatment of suture‐based VCD‐mediated closure in vivo and explain its mechanism with an ex vivo benchtop model.
Methods
This retrospective cohort study focused on patients who underwent LBAA closure complicated by vascular occlusion related to the Perclose VCD. These were managed through percutaneous endovascular interventions (PEI), in which the identified occlusions were treated using a peripheral cutting balloon. Additionally, a benchtop model utilizing silicone tubing was developed to simulate the occlusion mechanism caused by the Perclose VCD and to compare the use of non‐compliant versus cutting balloons in this context.
Results
A 6 mm cutting balloon successfully cut the monofilament polypropylene at 3 atmospheres (ATM), whereas a 5 mm non‐compliant balloon failed, even at a burst pressure of 24 ATM. In this patient cohort, seven patients were found to have common femoral artery (CFA) occlusion following the tightening of the previously deployed Perclose VCD. The median age was 87 years (interquartile range: 80−87), 71% were female, and had significant cardiovascular risk factors. All patients were diagnosed with CFA occlusion at the end of the transcatheter aortic valve replacement procedure during aortoiliac angiography, following vascular closure. PEI was performed using a combined cutting balloon and balloon angioplasty approach across the entire patient cohort with complete revascularization, minimal residual stenosis, no severe complications, no covered stents required, and no reported conversions to surgery. Long‐term follow‐up in six patients showed no reinterventions required or related mortalities.
Conclusion
Based on the findings from our ex vivo model and clinical experience, peripheral cutting balloon angioplasty should be considered for the management of vessel occlusion caused by suture‐based VCD.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, USA.
- Georgetown University School of Medicine, Washington, USA.
- Cardiovascular Intervention Program, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.
ORCID-Profile: Evan Czulada, Kalyan R Chitturi, Ilan Merdler, Ryan Wallace, Ron Waksman
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Haberman, D., Czulada, E., Lupu, L., Chitturi, K. R., Abusnina, W., Chaturvedi, A., Merdler, I., Galo, J., Verma, B. R., Wallace, R., Rogers, T., Ben-Dor, I., Waksman, R., & Bernardo, N. L. (2025). Percutaneous Management of Vessel Occlusion Caused by Suture‐Based Closure Devices: A Case Series and Benchtop Model. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31598
Vancouver
Haberman D, Czulada E, Lupu L, Chitturi KR, Abusnina W, Chaturvedi A, et al. Percutaneous Management of Vessel Occlusion Caused by Suture‐Based Closure Devices: A Case Series and Benchtop Model. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31598
BibTeX
@article{Haberman2025,
author = {Haberman, Dan and Czulada, Evan and Lupu, Lior and Chitturi, Kalyan R and Abusnina, Waiel and Chaturvedi, Abhishek and Merdler, Ilan and Galo, Jason and Verma, Beni R and Wallace, Ryan and Rogers, Toby and Ben-Dor, Itsik and Waksman, Ron and Bernardo, Nelson L},
title = {Percutaneous Management of Vessel Occlusion Caused by Suture‐Based Closure Devices: A Case Series and Benchtop Model},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31598}
}
EUVIMED