EUVIMEDEUVIMED entdecken
Fallbericht

Femoral Access‐Assisted Balloon Trapping: A Novel Approach to Managing Kinked Catheter During Transradial Coronary Intervention—A Case Report

Haşim Tüner iD, Fuat Polat iD, Özge Özden Kayhan iD, Ömer Göktekin iD

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31523 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Haşim Tüner und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Istanbul Arel University Memorial Bahcelievler Hospital, İstanbul, Türkiye. Hauptthemen laut MEDLINE: Angioplasty, Balloon, Coronary, Cardiac Catheterization, Cardiac Catheters, Catheterization, Peripheral. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 15 Literaturangaben. Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Catheter kinking is a rare but significant complication of trans‐radial coronary interventions, often caused by arterial tortuosity, vessel spasm, or improper catheter manipulation. We present a case of catheter kinking that occurred during coronary engagement in a transradial procedure, which could not be resolved using conventional techniques. Given the failure of guidewire advancement and deep inspiration maneuvers, a femoral access‐assisted balloon trapping technique was employed. This method allowed for the successful retrieval of the kinked catheter while minimizing vascular trauma and ensuring procedural completion. Catheter kinking during trans‐radial coronary interventions poses technical challenges requiring tailored management strategies. This case underscores the effectiveness of femoral access‐assisted balloon trapping technique as a viable and minimally invasive solution. Increased awareness and preparedness for such complications can enhance procedural success and patient safety.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Angioplasty, Balloon, CoronaryCardiac CatheterizationCardiac CathetersCatheterization, PeripheralDevice RemovalFemoral ArteryPercutaneous Coronary InterventionRadial ArteryAgedHumansMaleCoronary AngiographyEquipment DesignEquipment FailureTreatment Outcome

Stichwörter der Autor:innen

balloon trapping techniquecatheter kinkingtransradial accessvascular complications

Autor:innen und Institutionen

  • Department of Cardiology, Istanbul Arel University Memorial Bahcelievler Hospital, İstanbul, Türkiye.
  • Department of Cardiology, Health Science University Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Educatıon Research Hospıtal, İstanbul, Türkiye.
  • Sağlık Bilimleri Üniversitesi, Istanbul, Turkey.
  • Department of Cardiology, Memorial Bahçelievler Hospital, İstanbul, Türkiye.

ORCID-Profile: Haşim Tüner, Fuat Polat, Özge Özden Kayhan, Ömer Göktekin

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Tüner, H., Polat, F., Kayhan, Ö. Ö., & Göktekin, Ö. (2025). Femoral Access‐Assisted Balloon Trapping: A Novel Approach to Managing Kinked Catheter During Transradial Coronary Intervention—A Case Report. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31523

Vancouver

Tüner H, Polat F, Kayhan ÖÖ, Göktekin Ö. Femoral Access‐Assisted Balloon Trapping: A Novel Approach to Managing Kinked Catheter During Transradial Coronary Intervention—A Case Report. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31523

BibTeX
@article{Tner2025,
  author = {Tüner, Haşim and Polat, Fuat and Kayhan, Özge Özden and Göktekin, Ömer},
  title = {Femoral Access‐Assisted Balloon Trapping: A Novel Approach to Managing Kinked Catheter During Transradial Coronary Intervention—A Case Report},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31523}
}
Kostenlos starten →