Suture closure AFtEr large bore vein access (SAFE‐VEIN): A randomized, prospective study of the efficacy and safety of venous closure device
Catheterization and Cardiovascular Interventions · 2024
DOI: 10.1002/ccd.31173 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Catheterization and Cardiovascular Interventions, 2024. Verfasst von Mahmoud Ali und 18 weiteren Autor:innen. Beteiligte Einrichtung: Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Aurora Health Care, Milwaukee, Wisconsin, USA. Hauptthemen laut MEDLINE: Suture Techniques, Punctures, Hemostatic Techniques, Hemorrhage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Perclose ProGlide (PPG) Suture‐Mediated Closure System™ is safe and can reduce time to hemostasis following procedures requiring arterial access.
Aims
We aimed to compare PPG to figure of 8 suture in patients who underwent interventional catheter procedures requiring large bore venous access (LBVA) (≥13 French).
Methods
In this physician‐initiated, randomized, single‐center study [ clinicaltrials.gov ID: NCT04632641], single‐stick venous access was obtained under ultrasound guidance. Eligible patients were randomized 1:1, and 100 subjects received allocated treatment to either PPG ( n = 47) or figure of 8 suture ( n = 53). No femoral arterial access was used in any patient. Primary outcomes were time to achieve hemostasis (TTH) and time to ambulation (TTA). Secondary outcomes were time to discharge (TTD) and vascular‐related complications and mortality. Wilcoxon rank‐sum test was used to compare TTH, TTA, and TTD.
Results
TTH (minutes) was significantly lower in PPG versus figure of 8 suture [median, (Q1, Q3)] [7 (2,10) vs. 11 (10,15) respectively, p < 0.001]. TTA (minutes) was significantly lower in PPG compared to figure of 8 suture [322 (246,452) vs. 403 (353, 633) respectively, p = 0.005]. TTD (minutes) was not significantly different between the PPG and figure of 8 suture arms [1257 (1081, 1544) vs. 1338 (1171,1435), p = 0.650]. There was no difference in minor bleeding or access site hematomas between both arms. No other vascular complications or mortality were reported.
Conclusion
PPG use had lower TTH and TTA than figure of 8 suture in a population of patients receiving LBVA procedures. This may encourage same‐day discharge in these patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Aurora Health Care, Milwaukee, Wisconsin, USA.
- Academic Affairs, Cardiovascular Research, Aurora Sinai/Aurora St. Luke's Medical Centers, Aurora Health Care, Milwaukee, Wisconsin, USA.
- Academic Affairs, Aurora UW Medical Group, Center for Urban Population Health, Aurora Sinai Medical Center, Aurora Health Care, Milwaukee, Wisconsin, USA.
- Division of Cardiovascular Medicine, University of Wisconsin School of Medicine and Public Health, Milwaukee Clinical Campus, Milwaukee, Wisconsin, USA.
ORCID-Profile: Suhail Q Allaqaband, Tanvir Bajwa, Jodi Zilinski, M Eyman Mortada
Zitieren
APA
Ali, M., Masood, F., Erickson, L., Adefisoye, J., Kanani, J., Walczak, S., Ajam, T., Kieu, A., Premjee, M., Jan, M. F., Allaqaband, S. Q., Bajwa, T., Khitha, J., Zilinski, J., Jahangir, A., Djelmami-Hani, M., Sra, J., Niazi, I., & Mortada, M. E. (2024). Suture closure AFtEr large bore vein access (SAFE‐VEIN): A randomized, prospective study of the efficacy and safety of venous closure device. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31173
Vancouver
Ali M, Masood F, Erickson L, Adefisoye J, Kanani J, Walczak S, et al. Suture closure AFtEr large bore vein access (SAFE‐VEIN): A randomized, prospective study of the efficacy and safety of venous closure device. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31173
BibTeX
@article{Ali2024,
author = {Ali, Mahmoud and Masood, Faisal and Erickson, Lynn and Adefisoye, James and Kanani, Jim and Walczak, Sara and Ajam, Tarek and Kieu, Andy and Premjee, Mohammed and Jan, M Fuad and Allaqaband, Suhail Q and Bajwa, Tanvir and Khitha, Jayant and Zilinski, Jodi and Jahangir, Arshad and Djelmami-Hani, Mohamed and Sra, Jasbir and Niazi, Imran and Mortada, M Eyman},
title = {Suture closure AFtEr large bore vein access (SAFE‐VEIN): A randomized, prospective study of the efficacy and safety of venous closure device},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
doi = {10.1002/ccd.31173}
}
EUVIMED