Paclitaxel‐coated versus sirolimus‐coated balloon angioplasty for coronary artery disease: A systematic review and meta‐analysis
Catheterization and Cardiovascular Interventions · 2024-09 · 104 (3) · 425-436
DOI: 10.1002/ccd.31154 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 24.07.2024. Verfasst von Doosup Shin und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, St Francis Hospital and Heart Center, Roslyn, New York, USA. Hauptthemen laut MEDLINE: Angioplasty, Balloon, Coronary, Cardiac Catheters, Cardiovascular Agents, Coated Materials, Biocompatible. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 36 Literaturangaben.
Abstract (Originalsprache)
Background
Although use of sirolimus‐based analogs has shown superiority over paclitaxel in drug‐eluting stents, the relative efficacy of these two agents released from drug‐coated balloons (DCB) is unclear. The present meta‐analysis is aimed to compare outcomes after percutaneous coronary intervention (PCI) with paclitaxel‐coated balloons (PCB) versus sirolimus‐coated balloons (SCB) for either in‐stent restenosis or native de novo lesions.
Methods
The study outcomes were 1) target lesion failure (TLF), a composite of cardiac death, target vessel myocardial infarction, or target lesion revascularization, and 2) follow‐up angiographic parameters including late lumen loss (LLL), diameter stenosis, and minimal lumen diameter (MLD). Pooled odds ratios (OR) and weighted mean differences (WMD) with 95% confidence intervals (CI) were calculated by using random‐effects models.
Results
A search of PubMed, EMBASE, and Cochrane Library from their inception to January 2024 identified five randomized clinical trials and three observational studies with a total of 1861 patients (889 in PCB and 972 in SCB groups). During 9–12 months of follow‐up, there was no significant difference in TLF (OR 1.01, 95% CI 0.75–1.35) between the two groups. On follow‐up angiography at 6–9 months, MLD (WMD 0.10, 95% CI 0.02–0.17) was larger in PCB but there was no statistically significant difference in LLL (WMD −0.11, 95% CI −0.23–0.02) and diameter stenosis (WMD −3.33, 95% CI −8.11–1.45).
Conclusions
Among patients undergoing DCB‐only PCI, the risk of TLF was similar during 9–12 months of follow‐up after PCB and SCB treatment. However, the MLD was larger favoring PCB over SCB on follow‐up angiography.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiology, St Francis Hospital and Heart Center, Roslyn, New York, USA.
- Clinical and Experimental Interventional Cardiology, University of Saarland, Homburg/Saar, Germany.
- Division of Cardiology, Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, New York, USA.
- Piedmont Heart Institute, Atlanta, Georgia, USA.
- Department of Cardiology, Freeman Hospital and Newcastle University, Newcastle upon Tyne, UK.
- Department of Cardiology, University Hospital Galway, Galway, Ireland.
ORCID-Profile: Doosup Shin, Mandeep Singh, Evan Shlofmitz, Bruno Scheller, Azeem Latib, David E Kandzari
Zitieren
APA
Shin, D., Singh, M., Shlofmitz, E., Scheller, B., Latib, A., Kandzari, D. E., Zaman, A., Mylotte, D., Dakroub, A., Malik, S., Sakai, K., Jeremias, A., Moses, J. W., Shlofmitz, R. A., Stone, G. W., & Ali, Z. A. (2024). Paclitaxel‐coated versus sirolimus‐coated balloon angioplasty for coronary artery disease: A systematic review and meta‐analysis. Catheterization and Cardiovascular Interventions, 104(3), 425-436. https://doi.org/10.1002/ccd.31154
Vancouver
Shin D, Singh M, Shlofmitz E, Scheller B, Latib A, Kandzari DE, et al. Paclitaxel‐coated versus sirolimus‐coated balloon angioplasty for coronary artery disease: A systematic review and meta‐analysis. Catheterization and Cardiovascular Interventions. 2024;104(3):425-436. doi:10.1002/ccd.31154
BibTeX
@article{Shin2024,
author = {Shin, Doosup and Singh, Mandeep and Shlofmitz, Evan and Scheller, Bruno and Latib, Azeem and Kandzari, David E and Zaman, Azfar and Mylotte, Darren and Dakroub, Ali and Malik, Sarah and Sakai, Koshiro and Jeremias, Allen and Moses, Jeffrey W and Shlofmitz, Richard A and Stone, Gregg W and Ali, Ziad A},
title = {Paclitaxel‐coated versus sirolimus‐coated balloon angioplasty for coronary artery disease: A systematic review and meta‐analysis},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
volume = {104},
number = {3},
pages = {425--436},
doi = {10.1002/ccd.31154}
}