EUVIMEDEUVIMED entdecken
Crossref

Chromatographic Method Validation for Simultaneous Determination of Bergenin and Alpha‐Lipoic Acid in Skin Penetration Studies

Samuel Dutra-Portes, Clara S Alves, Breno N Matos, Tais Gratieri iD, Marcilio Cunha-Filho iD, Emerson S Lima iD, Guilherme M Gelfuso iD

Biomedical Chromatography · 2026-08 · 40 (8)

DOI: 10.1002/bmc.70538 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Chromatography, online erschienen am 29.06.2026. Verfasst von Samuel Dutra-Portes und 6 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Food, Drugs, and Cosmetics (LTMAC), University of Brasilia, Brasília, Distrito Federal, Brazil. Hauptthemen laut MEDLINE: Benzopyrans, Skin, Thioctic Acid, Skin Absorption. Die Arbeit nennt 39 Literaturangaben. Gefördert von: Conselho Nacional de Desenvolvimento Científico e Tecnológico, University of Brasília, Universidade de Brasília.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study validates a selective, robust high‐performance liquid chromatography (HPLC) method for the simultaneous determination of bergenin (BER) and alpha‐lipoic acid (ALA) in skin permeation studies. Separation used a Discovery C 18 column (250 × 4.6 mm, 5 μm) and a mobile phase of acetonitrile and aqueous phosphoric acid (pH 3.0). A multistep gradient (10%–30% acetonitrile from 0–2 min; 30%–40% from 2–6 min; and 40%–60% from 6–9 min) at 1.0 mL/min enabled a 20‐min run time with UV detection at 210 nm. Following ICH M10 bioanalytical guidelines, the method showed linearity ( r > 0.999) from 0.5 to 15 μg/mL for both analytes. Limits of detection were 0.120 and 0.083 μg/mL, and limits of quantification were 0.363 and 0.250 μg/mL for BER and ALA, respectively. Precision and accuracy were satisfactory, with a relative standard deviation (RSD) < 3.5%, while recoveries ranged between 80.8% and 96.2% (RSD < 10.0%) from porcine stratum corneum and viable skin. The method proved robust against variations in temperature, flow rate, and column batch, without endogenous skin interference. In conclusion, this validated method is a reliable tool for supporting the development of topical formulations containing the association of BER and ALA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BenzopyransSkinThioctic AcidSkin AbsorptionAnimalsSwineReproducibility of ResultsChromatography, High Pressure LiquidLimit of DetectionLinear Models

Stichwörter der Autor:innen

antioxidantsliquid chromatographyskin permeationtopical formulation

Substanzen

BenzopyransbergeninThioctic Acid

Autor:innen und Institutionen

  • Laboratory of Food, Drugs, and Cosmetics (LTMAC), University of Brasilia, Brasília, Distrito Federal, Brazil.
  • Faculty of Pharmaceutical Sciences, Federal University of Amazonas, Manaus, Amazonas, Brazil.

ORCID-Profile: Tais Gratieri, Marcilio Cunha-Filho, Emerson S Lima, Guilherme M Gelfuso

Förderung

  • Conselho Nacional de Desenvolvimento Científico e Tecnológico
  • University of Brasília
  • Universidade de Brasília

Zitieren

APA

Dutra-Portes, S., Alves, C. S., Matos, B. N., Gratieri, T., Cunha-Filho, M., Lima, E. S., & Gelfuso, G. M. (2026). Chromatographic Method Validation for Simultaneous Determination of Bergenin and Alpha‐Lipoic Acid in Skin Penetration Studies. Biomedical Chromatography, 40(8). https://doi.org/10.1002/bmc.70538

Vancouver

Dutra-Portes S, Alves CS, Matos BN, Gratieri T, Cunha-Filho M, Lima ES, et al. Chromatographic Method Validation for Simultaneous Determination of Bergenin and Alpha‐Lipoic Acid in Skin Penetration Studies. Biomedical Chromatography. 2026;40(8). doi:10.1002/bmc.70538

BibTeX
@article{DutraPortes2026,
  author = {Dutra-Portes, Samuel and Alves, Clara S and Matos, Breno N and Gratieri, Tais and Cunha-Filho, Marcilio and Lima, Emerson S and Gelfuso, Guilherme M},
  title = {Chromatographic Method Validation for Simultaneous Determination of Bergenin and Alpha‐Lipoic Acid in Skin Penetration Studies},
  journal = {Biomedical Chromatography},
  year = {2026},
  volume = {40},
  number = {8},
  doi = {10.1002/bmc.70538}
}
Kostenlos starten →