Development and Validation of a Stability‐Indicating Reversed‐Phase Liquid Chromatographic Method for the Simultaneous Determination of Methylprednisolone Aceponate and Benzyl Alcohol in Cream Formulation
Biomedical Chromatography · 2026-05 · 40 (5)
DOI: 10.1002/bmc.70429 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Chromatography, online erschienen am 20.03.2026. Verfasst von Aysen Kurt Cucu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Analytical Chemistry, Faculty of Pharmacy, Marmara University, Istanbul, Turkey. Hauptthemen laut MEDLINE: Chromatography, Reverse-Phase, Benzyl Alcohol, Methylprednisolone. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 48 Literaturangaben. Gefördert von: Marmara University Scientific Research Projects Commission.
Abstract (Originalsprache)
Methylprednisolone aceponate (MPA) is a nonhalogenated, diester corticosteroid, specifically designed to treat eczema. The amount of benzyl alcohol (BA), an excipient, is quite high in the cream formulation of MPA; therefore, the simultaneous determination of MPA, active ingredient, and BA is important. Chromatographic analysis was performed on an Agilent Poroshell 120 EC‐C18 (100 × 2.1 mm 2.7 μm) column using acetonitrile–water (40:60, v/v) mixture as mobile phase with isocratic elution. The diode array detector (DAD) was set at 245 and 258 nm. The column was thermostated at 40°C ± 1°C. A flow rate of 0.5 mL/min was used to elute the analyte from the column. The calibration curve was linear in the concentration range 0.17–4.20 μg/mL for MPA and 1.50–35.0 μg/mL for BA. The R 2 values for MPA and BA were found to be 0.9989 and 0.9999, respectively. The limit of detection was found to be 0.048 μg/mL for MPA, 0.46 μg/mL for BA, and the limit of quantification was found to be 0.17 μg/mL for MPA and 1.50 μg/mL for BA. MPA was exposed to stress conditions, and the stressed samples were analyzed using the proposed method, which can effectively separate MPA from its degradation products.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Analytical Chemistry, Faculty of Pharmacy, Marmara University, Istanbul, Turkey.
- Department of Analytical Chemistry, Institute of Health Science, Marmara University, Istanbul, Turkey.
- Department of Analytical Chemistry, Faculty of Pharmacy, Istanbul University, Istanbul, Turkey.
ORCID-Profile: Aysen Kurt Cucu, Deniz Çıkla Yılmaz, Fevziye Gizem Tecimen, Aykut Kul
Förderung
- Marmara University Scientific Research Projects Commission
Zitieren
APA
Cucu, A. K., Yılmaz, D. Ç., Tecimen, F. G., & Kul, A. (2026). Development and Validation of a Stability‐Indicating Reversed‐Phase Liquid Chromatographic Method for the Simultaneous Determination of Methylprednisolone Aceponate and Benzyl Alcohol in Cream Formulation. Biomedical Chromatography, 40(5). https://doi.org/10.1002/bmc.70429
Vancouver
Cucu AK, Yılmaz DÇ, Tecimen FG, Kul A. Development and Validation of a Stability‐Indicating Reversed‐Phase Liquid Chromatographic Method for the Simultaneous Determination of Methylprednisolone Aceponate and Benzyl Alcohol in Cream Formulation. Biomedical Chromatography. 2026;40(5). doi:10.1002/bmc.70429
BibTeX
@article{Cucu2026,
author = {Cucu, Aysen Kurt and Yılmaz, Deniz Çıkla and Tecimen, Fevziye Gizem and Kul, Aykut},
title = {Development and Validation of a Stability‐Indicating Reversed‐Phase Liquid Chromatographic Method for the Simultaneous Determination of Methylprednisolone Aceponate and Benzyl Alcohol in Cream Formulation},
journal = {Biomedical Chromatography},
year = {2026},
volume = {40},
number = {5},
doi = {10.1002/bmc.70429}
}
EUVIMED