Development and validation of a stability‐indicating ultra‐high‐performance liquid chromatography method for the estimation of ibrutinib and trace‐level quantification of related substances using quality‐by‐design approach
Biomedical Chromatography · 2024-03 · 38 (3)
DOI: 10.1002/bmc.5798 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Chromatography, online erschienen am 11.12.2023. Verfasst von Shishir Kumar Prasad und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmaceutical Chemistry, College of Pharmaceutical Sciences, Dayananda Sagar University, Bengaluru, India. Hauptthemen laut MEDLINE: Piperidines, Formates. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
A new ultra‐high‐performance liquid chromatography method was developed using quality‐by‐design principles for quantifying trace‐level impurities of ibrutinib. The method utilized an ACQUITY UPLC BEH C18 column with a mobile phase consisting of equal parts of 0.02 M formic acid in water and 0.02 M formic acid in acetonitrile. The critical method parameters, including mobile phase pH, column temperature, and flow rate, were optimized using the design of experiments. Statistical analysis revealed the impact of these parameters on critical quality attributes. Perturbation and response surface plots illustrated the individual and interactive effects of the parameters. The optimal parameter levels were determined to be pH, 2.5; column temperature, 28°C; and flow rate, 0.55 mL/min. Confirmation experiments demonstrated the method's robustness, with the separation of impurities and unknown degradation products within a 5‐min runtime. The optimized ultra‐performance liquid chromatography method was validated according to ICH guidelines. The method exhibited linear response within the range of 0.025–100 μg/mL for ibrutinib and 0.0187–0.225 μg/mL for impurities ( r 2 > 0.9995), with limits of detection/limits of quantification of 0.01/0.025 and 0.015/0.0187 for ibrutinib and four impurities, respectively. Recoveries for the drug and impurities ranged from 92.69 to 102.7%, and precision was below 2% and 8% relative standard deviation for ibrutinib and impurities, respectively.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Pharmaceutical Chemistry, College of Pharmaceutical Sciences, Dayananda Sagar University, Bengaluru, India.
ORCID-Profile: Divekar Kalpana
Zitieren
APA
Prasad, S. K., & Kalpana, D. (2024). Development and validation of a stability‐indicating ultra‐high‐performance liquid chromatography method for the estimation of ibrutinib and trace‐level quantification of related substances using quality‐by‐design approach. Biomedical Chromatography, 38(3). https://doi.org/10.1002/bmc.5798
Vancouver
Prasad SK, Kalpana D. Development and validation of a stability‐indicating ultra‐high‐performance liquid chromatography method for the estimation of ibrutinib and trace‐level quantification of related substances using quality‐by‐design approach. Biomedical Chromatography. 2024;38(3). doi:10.1002/bmc.5798
BibTeX
@article{Prasad2024,
author = {Prasad, Shishir Kumar and Kalpana, Divekar},
title = {Development and validation of a stability‐indicating ultra‐high‐performance liquid chromatography method for the estimation of ibrutinib and trace‐level quantification of related substances using quality‐by‐design approach},
journal = {Biomedical Chromatography},
year = {2024},
volume = {38},
number = {3},
doi = {10.1002/bmc.5798}
}