Vergleichsstudie in der Zeitschrift Biotechnology and Bioengineering, 2025-11. Verfasst von Maarten Klaverdijk und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biotechnology, Delft University of Technology, Delft, The Netherlands. Hauptthemen laut MEDLINE: Spectrum Analysis, Raman, Bioreactors, Saccharomyces cerevisiae, Models, Biological. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: This project is funded by the Department of Biotechnology at Delft University of Technology..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In‐line Raman spectroscopy combined with accurate quantification models can offer detailed real‐time insights into a bioprocess by monitoring key process parameters. However, traditional approaches for model calibration require extensive data collection from multiple bioreactor runs, resulting in process‐specific models that are sensitive to operational changes. These challenges can be tackled by simplifying experimental data generation or implementation of computational methods to obtain synthetic and augmented Raman spectra. In this study, we utilized a small experimental dataset of 16 single compound spectra to calibrate quantification models by using partial least squares (PLS) and indirect hard modeling (IHM), leading to comparable rRMSEP values for glucose (4.8% and 4.2%), ethanol (11.6% and 6.3%), and biomass (16.2% and 10.0%) when applied to yeast batch and fed‐batch bioprocesses. Subsequently, isolated spectral features extracted during IHM were used to generate fully synthetic spectral datasets for PLS model calibration, resulting in rRMSEPs of 3.2% and 14.5% for glucose and ethanol, respectively. Finally, spectra from a single batch process were augmented with the same isolated spectral features, and calibration with these augmented spectra reduced rRMSEP by 18.6% point (glucose) and 4.3% point (ethanol) compared to process‐only calibrated models. This study demonstrates how different approaches may support robust development and rapid implementation of Raman spectroscopy‐based models while minimizing experimental efforts, where even complete independence of process data can be achieved.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spectrum Analysis, RamanBioreactorsSaccharomyces cerevisiaeModels, BiologicalCalibrationGlucoseEthanolBiomassBatch Cell Culture Techniques

Stichwörter der Autor:innen

Raman spectroscopySaccharomyces cerevisiaebioprocessingdata augmentationin‐line measurementsprocess analytical technology

Substanzen

GlucoseEthanol

Autor:innen und Institutionen

  • Department of Biotechnology, Delft University of Technology, Delft, The Netherlands.

ORCID-Profile: Maarten Klaverdijk

Förderung

  • This project is funded by the Department of Biotechnology at Delft University of Technology.

Zitieren

APA

Klaverdijk, M., Smulders, L. A., Ottens, M., & Klijn, M. E. (2026). Towards Rapid Calibration of Bioprocess Quantification Models Using Single Compound Raman Spectra: A Comparison of Four Approaches. Biotechnology and Bioengineering, 123(2), 324-336. https://doi.org/10.1002/bit.70092

Vancouver

Klaverdijk M, Smulders LA, Ottens M, Klijn ME. Towards Rapid Calibration of Bioprocess Quantification Models Using Single Compound Raman Spectra: A Comparison of Four Approaches. Biotechnology and Bioengineering. 2026;123(2):324-336. doi:10.1002/bit.70092

BibTeX
@article{Klaverdijk2026,
  author = {Klaverdijk, Maarten and Smulders, Lisa A and Ottens, Marcel and Klijn, Marieke E},
  title = {Towards Rapid Calibration of Bioprocess Quantification Models Using Single Compound Raman Spectra: A Comparison of Four Approaches},
  journal = {Biotechnology and Bioengineering},
  year = {2026},
  volume = {123},
  number = {2},
  pages = {324--336},
  doi = {10.1002/bit.70092}
}
Kostenlos starten →