Comparison of p‐Tm:YAG, TFL and Ho:YAG's in vitro ablation rates on synthetic and human stones
BJUI Compass · 2025-08 · 6 (8)
DOI: 10.1002/bco2.70067 ↗Kurz gefasst
Fachartikel in der Zeitschrift BJUI Compass, online erschienen am 18.08.2025. Verfasst von Frédéric Panthier und 13 weiteren Autor:innen. Beteiligte Einrichtung: Sorbonne University GRC Urolithiasis no. 20, Tenon Hospital Paris France. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Association of Urology, Association Française d'Urologie.
Abstract (Originalsprache)
Objective
To compare in vitro the ablation rates of p‐Tm:YAG, TFL and Ho:YAG against synthetic and human stones. Material and Methods p‐Tm:YAG, TFL and Low‐Power (LP) Ho:YAG were compared using 270 μm core‐diameter laser fibres (CDF); experiments with 200 μm(p‐Tm:YAG) and 150 μm‐CDF (TFL) were also included. A continuous laser emission was applied through a spiral trajectory for 20 seconds with the laser fibre tip in contact with synthetic hard (HSP) and soft stone phantoms (SSP) submerged in saline. “Dusting” settings for p‐Tm:YAG(0,6 J‐20 Hz‐Flex Long Pulse), TFL(0,5 J‐30 Hz‐Short Pulse) and Ho:YAG(0,5 J‐30 Hz‐Long Pulse) and “Fragmentation” settings for p‐Tm:YAG(1 J‐15 Hz‐Captive), TFL(1 J‐15 Hz‐Short Pulse) and Ho:YAG(1 J‐15 Hz‐Long Pulse) were analysed. Then, experiments for human calcium oxalate monohydrate (COM), uric acid (UA) and cystine (CYS) stones were performed with single laser pulses at 0.6 J, 0.8 J and 1.0 J for p‐Tm:YAG (Captive Fragmenting mode), TFL (Short Pulse) and Ho:YAG (Long Pulse). Synthetic and human stone samples were dried before three‐dimensional scanning to measure ablation rates (ARs) and ablation volume per pulse (AVP).
Results
For synthetic stones with 270 μm‐CDF, the p‐Tm:YAG and TFL presented similar ARs, except in Fragmentation against HSP (95,1 ± 13,6vs67 ± 14 p = 0,02, respectively). Both p‐Tm:YAG and TFL achieved higher ARs than Ho:YAG in all settings. p‐Tm:YAG‐200 μm‐CDF and TFL‐150 μm‐CDF presented similar ARs, except in Fragmentation against HSP(78,4 ± 8vs42,5 ± 2,6 mm 3 /min,p = 0,0002). Both p‐Tm:YAG‐200 μm‐CDF and TFL‐150 μm‐CDF presented at least 50% higher ARs than 270 μm‐Ho:YAG. For human stones with COM, TFL exhibited higher AVP compared to p‐Tm:YAG and Ho:YAG across all pulse energies (258,2 ± 213vs81,7 ± 31,9vs41,5 ± 25,4 μm 3 p = 0,01, respectively). Against UA, Ho:YAG demonstrated higher AVP compared to TFL and p‐Tm:YAG (355,2 ± 161vs99,8 ± 76,7vs292,9 ± 203,1 μm 3 p = 0,0005, respectively). For CYS, Ho:YAG presented higher AVP but without significance (99,8 ± 76,7 vs 49,3 ± 36,3 vs 38,8 ± 12,2 μm 3 , p = 0,09). Conclusion p‐Tm:YAG and TFL achieved higher ARs than LP‐Ho:YAG against synthetic stones in vitro. For human stones, TFL achieved the highest AVP against COM while LP‐Ho:YAG delivered higher AVPs against UA and CYS, for which TFL performed worst.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Sorbonne University GRC Urolithiasis no. 20, Tenon Hospital Paris France.
- PIMM UMR 8006 CNRS-Arts et Métiers ParisTech Paris France.
- Progressive Endourological Association for Research and Leading Solutions (PEARLS) Paris France.
- Department of Urology Westmoreland Street Hospital, UCLH NHS Foundation Trust London UK.
- Endourology Technology Section of European Association of Urology (EAU) Arnhem The Netherlands.
- Department of Urology Hospital Clinic -University of Barcelona Barcelona Spain.
ORCID-Profile: Frédéric Panthier, Alba Sierra, Etienne Xavier Keller, Eugenio Ventimiglia, Jia-Lun Kwok, Mariela Corrales, Daron Smith
Förderung
- European Association of Urology
- Association Française d'Urologie
Zitieren
APA
Panthier, F., Sierra, A., Keller, E. X., Chicaud, M., Ventimiglia, E., Kwok, J. L., De Coninck, V., Corrales, M., Daudon, M., Gorny, C., Doizi, S., Berthe, L., Smith, D., & Traxer, O. (2025). Comparison of p‐Tm:YAG, TFL and Ho:YAG's in vitro ablation rates on synthetic and human stones. BJUI Compass, 6(8). https://doi.org/10.1002/bco2.70067
Vancouver
Panthier F, Sierra A, Keller EX, Chicaud M, Ventimiglia E, Kwok JL, et al. Comparison of p‐Tm:YAG, TFL and Ho:YAG's in vitro ablation rates on synthetic and human stones. BJUI Compass. 2025;6(8). doi:10.1002/bco2.70067
BibTeX
@article{Panthier2025,
author = {Panthier, Frédéric and Sierra, Alba and Keller, Etienne Xavier and Chicaud, Marie and Ventimiglia, Eugenio and Kwok, Jia-Lun and De Coninck, Vincent and Corrales, Mariela and Daudon, Michel and Gorny, Cyril and Doizi, Steeve and Berthe, Laurent and Smith, Daron and Traxer, Olivier},
title = {Comparison of p‐Tm:YAG, TFL and Ho:YAG's in vitro ablation rates on synthetic and human stones},
journal = {BJUI Compass},
year = {2025},
volume = {6},
number = {8},
doi = {10.1002/bco2.70067}
}
EUVIMED