Ru(bpy) 3 2+ ‐DBAE Coreactant‐SDS Enhancer Systems on Automated Electrochemiluminescence Immunoassay Analyzer Containing Pt Electrode for Highly Sensitive Detection of Carbohydrate Antigen 19‐9
Chemistry – An Asian Journal · 2026
DOI: 10.1002/asia.202500998 ↗Kurz gefasst
Fachartikel in der Zeitschrift Chemistry – An Asian Journal, 2026. Verfasst von Xiaolin Yang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Analytical Chemistry For Life Science of Shaanxi Province, School of Chemistry and Chemical Engineering, Shaanxi Normal University,…. Hauptthemen laut MEDLINE: Electrochemical Techniques, Sodium Dodecyl Sulfate, CA-19-9 Antigen, Luminescent Measurements. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Fundamental Research Funds for the Central Universities.
Abstract (Originalsprache)
This work reports one Ru(bpy) 3 2+ ‐N,N‐dibutylethanolamine (DBAE) coreactant‐sodium dodecyl sulfate (SDS) enhancer system on an automated electrochemiluminescence immunoassay (ECLIA) analyzer for the ECL immunoassay of carbohydrate antigen 19‐9 (CA19‐9). The ECL intensity of the MB@SA•biotin‐GSK‐Ru1 bio‐conjugates in DBAE solution on the automated ECLIA analyzer, which employs a platinum working electrode under the “Heterogeneous” model, demonstrates a 333% enhancement in comparison with that of the conventional tri‐n‐propylamine coreactant. Subsequently, the developed novel DBAE coreactant‐based automated ECL method was estimated by the determination of CA19‐9, while the MB@SA•biotin‐Ab1 was taken as the capture probe and the Ru(bpy) 3 2+ derivative‐labeled CA19‐9 antibody (Ab2‐Ru1) was employed as signal probe. In the presence of the SDS enhancer, the relative ECL intensity of the formed “sandwich” MB@SA•biotin‐Ab1/CA19‐9/Ab2‐Ru1 biocomplexes exhibits a 4.3‐fold enhancement on the automated ECLIA analyzer. Notably, under the optimized conditions, the increased ECL intensity was directly in proportion to the concentration of CA19‐9 in the range from 0.5 to 8.0 U/mL with a limit of detection of 0.25 U/mL. This approach has the potential to be extended to magnetic bead‐based automated ECL methods for the sensitive detection of clinical protein biomarkers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Key Laboratory of Analytical Chemistry For Life Science of Shaanxi Province, School of Chemistry and Chemical Engineering, Shaanxi Normal University, Xi'an, P.R. China.
- Department of Central Laboratory, Xi'an No.1 Hospital, The First Affiliated Hospital of Northwest University, Xi'an, P.R. China.
ORCID-Profile: Xiaolin Yang, Honglan Qi, Chengxiao Zhang
Förderung
- National Natural Science Foundation of China
- Fundamental Research Funds for the Central Universities
Zitieren
APA
Yang, X., Wang, Z., Wang, C., Ma, Y., Wei, Y., Qi, H., & Zhang, C. (2026). Ru(bpy) 3 2+ ‐DBAE Coreactant‐SDS Enhancer Systems on Automated Electrochemiluminescence Immunoassay Analyzer Containing Pt Electrode for Highly Sensitive Detection of Carbohydrate Antigen 19‐9. Chemistry – An Asian Journal. https://doi.org/10.1002/asia.202500998
Vancouver
Yang X, Wang Z, Wang C, Ma Y, Wei Y, Qi H, et al. Ru(bpy) 3 2+ ‐DBAE Coreactant‐SDS Enhancer Systems on Automated Electrochemiluminescence Immunoassay Analyzer Containing Pt Electrode for Highly Sensitive Detection of Carbohydrate Antigen 19‐9. Chemistry – An Asian Journal. 2026. doi:10.1002/asia.202500998
BibTeX
@article{Yang2026,
author = {Yang, Xiaolin and Wang, Zimei and Wang, Chengkai and Ma, Yaping and Wei, Yuxi and Qi, Honglan and Zhang, Chengxiao},
title = {Ru(bpy) 3 2+ ‐DBAE Coreactant‐SDS Enhancer Systems on Automated Electrochemiluminescence Immunoassay Analyzer Containing Pt Electrode for Highly Sensitive Detection of Carbohydrate Antigen 19‐9},
journal = {Chemistry – An Asian Journal},
year = {2026},
doi = {10.1002/asia.202500998}
}
EUVIMED