EUVIMEDEUVIMED entdecken
Open Access

Homemade Ultrasound Stand‐Off Gel Pads: A Multi‐Parameter Performance Study

Muhammad Fiqqrie De Gracious Bin Mohd Firdaus William, Nur Arisha Binti Jamil, Sook Sam Leong iD, Rafidah Supar, Nazimah Ab Mumin, Ahmad Razali Ishak, Nurul Syafiqah Binti Noor Idzam, Mohd Nazzihan Md Ajis, Anushya Vijayananthan

Australasian Journal of Ultrasound in Medicine · 2026-05 · 29 (2)

DOI: 10.1002/ajum.70044 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Australasian Journal of Ultrasound in Medicine, online erschienen am 21.05.2026. Verfasst von Muhammad Fiqqrie De Gracious Bin Mohd Firdaus William und 8 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Imaging Studies Universiti Teknologi MARA Selangor Malaysia. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

Stand‐off gel pads made of natural ingredients may provide a low‐cost option in clinical ultrasound imaging. This study investigated the performance and safety of homemade pads.

Methods

Thirty stand‐off gel pads were made of aloe vera, xanthan gum and cucumber. Testing of each was performed to assess mechanical failure, moisture loss, contamination rate and image quality. Mechanical strength was measured using a tensile testing machine; drying rate was calculated using the moisture loss formula; contamination was evaluated through colony counting; and image quality was assessed using a chicken breast phantom.

Results

There was a significant difference in compressive force resistance between the stand‐off gel pads ( p = 0.044). The force required was higher for those made of xanthan gum compared with cucumber ( p = 0.047). There was a significant difference in moisture loss ( p < 0.001). Aloe vera gel exhibited a higher moisture loss (9.65% ± 0.15%) compared with xanthan gum (5.94% ± 0.11%) and cucumber (5.81% ± 0.11%) ( p < 0.001). There was a significant difference in axial resolution ( p = 0.037), lateral resolution ( p = 0.029) and striations in the chicken breast parenchyma ( p = 0.037). Speckle artefacts are presented in xanthan‐gum‐based stand‐off gel pad. There was a borderline significant difference of artefact presence in images produced using all pads ( p = 0.050). Microbial counts were too low to be quantified.

Conclusion

A cucumber stand‐off gel pad is a practical alternative for clinical use owing to its structural integrity, retained moisture and low microbial contamination. The image quality produced using the pad was comparable with the conventional glove‐filled‐with‐water stand‐off method.

Schlagwörter

Stichwörter der Autor:innen

contamination ratehomemadeimage qualitymechanical failuremoisture lossstand‐off gel pad

Autor:innen und Institutionen

  • Centre for Medical Imaging Studies Universiti Teknologi MARA Selangor Malaysia.
  • M&S Medica Supplies Sdn Bhd Johor Malaysia.
  • Medical Imaging & Radiomics AI Lab (MIRA Lab) Universiti Teknologi MARA Selangor Malaysia.
  • Faculty of Medicine Universiti Teknologi MARA Selangor Malaysia.
  • Centre for Environmental Health & Safety Studies Universiti Teknologi MARA Selangor Malaysia.
  • Faculty of Medicine Universiti of Malaya Kuala Lumpur Malaysia.

ORCID-Profile: Sook Sam Leong

Zitieren

APA

William, M. F. G. B. M. F., Jamil, N. A. B., Leong, S. S., Supar, R., Ab Mumin, N., Ishak, A. R., Idzam, N. S. B. N., Ajis, M. N. M., & Vijayananthan, A. (2026). Homemade Ultrasound Stand‐Off Gel Pads: A Multi‐Parameter Performance Study. Australasian Journal of Ultrasound in Medicine, 29(2). https://doi.org/10.1002/ajum.70044

Vancouver

William MFGBMF, Jamil NAB, Leong SS, Supar R, Ab Mumin N, Ishak AR, et al. Homemade Ultrasound Stand‐Off Gel Pads: A Multi‐Parameter Performance Study. Australasian Journal of Ultrasound in Medicine. 2026;29(2). doi:10.1002/ajum.70044

BibTeX
@article{William2026,
  author = {William, Muhammad Fiqqrie De Gracious Bin Mohd Firdaus and Jamil, Nur Arisha Binti and Leong, Sook Sam and Supar, Rafidah and Ab Mumin, Nazimah and Ishak, Ahmad Razali and Idzam, Nurul Syafiqah Binti Noor and Ajis, Mohd Nazzihan Md and Vijayananthan, Anushya},
  title = {Homemade Ultrasound Stand‐Off Gel Pads: A Multi‐Parameter Performance Study},
  journal = {Australasian Journal of Ultrasound in Medicine},
  year = {2026},
  volume = {29},
  number = {2},
  doi = {10.1002/ajum.70044}
}