Wirelessly Actuated Ciliary Airway Stent for Excessive Mucus Transportation
Advanced Materials Technologies · 2023-12 · 8 (23)
DOI: 10.1002/admt.202301003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Advanced Materials Technologies, online erschienen am 31.10.2023. Verfasst von Yusheng Wang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Vanderbilt University, Nashville, TN 37240, USA. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS.
Abstract (Originalsprache)
Small‐scale cilia‐like devices that can manipulate fluids in narrow spaces have great potential in microfluidics, biomechanics, biomedical engineering, and other applications. However, prior studies mostly focus on artificial cilia for pumping fluids in lab‐on‐a‐chip microfluidic applications. The design and control of artificial cilia for transporting viscous mucus in confined and tubular structures remain challenging and medical devices such as airway stents with ciliary function are still missing. Herein, a method is reported that enables integrating artificial cilia arrays on 3D curved surfaces and an airway ciliary stent is presented for excessive mucus transportation. The method allows encoding bioinspired non‐reciprocal motion and metachronal waves for efficient fluid pumping in tubular structures. The method also introduces a lubricant hydrogel coating layer on artificial cilia inspired by the periciliary layer in airway cilia, which further enhances viscous fluid transportation. It is demonstrated that a novel ciliary airway stent can transport viscous porcine mucus in a lung phantom even faster than the respiratory cilia in a healthy human lung. The methods of designing, integrating, and controlling artificial cilia on 3D curved surfaces thus enable the unprecedented function of removing excessive mucus beyond traditional airway stents for treating various lung diseases in a minimally invasive manner.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Vanderbilt University, Nashville, TN 37240, USA.
- Vanderbilt Institute for Surgery and Engineering, Vanderbilt University, Nashville, TN 37240, USA.
- School of Medicine Vanderbilt University, Nashville, TN 37240, USA.
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN 37240,USA.
Förderung
- NIBIB NIH HHS
Zitieren
APA
Wang, Y., Sharma, S., Maldonado, F., & Dong, X. (2023). Wirelessly Actuated Ciliary Airway Stent for Excessive Mucus Transportation. Advanced Materials Technologies, 8(23). https://doi.org/10.1002/admt.202301003
Vancouver
Wang Y, Sharma S, Maldonado F, Dong X. Wirelessly Actuated Ciliary Airway Stent for Excessive Mucus Transportation. Advanced Materials Technologies. 2023;8(23). doi:10.1002/admt.202301003
BibTeX
@article{Wang2023,
author = {Wang, Yusheng and Sharma, Saksham and Maldonado, Fabien and Dong, Xiaoguang},
title = {Wirelessly Actuated Ciliary Airway Stent for Excessive Mucus Transportation},
journal = {Advanced Materials Technologies},
year = {2023},
volume = {8},
number = {23},
doi = {10.1002/admt.202301003}
}
EUVIMED