A Multi‐Channel Handheld Diagnostic Device Based on Green Biomimetic Material for Point‐of‐Care Testing of Vitamin C in Human Fluids
Advanced Healthcare Materials · 2025
DOI: 10.1002/adhm.202403454 ↗Kurz gefasst
Fachartikel in der Zeitschrift Advanced Healthcare Materials, 2025. Verfasst von Donglei Fu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Hubei Engineering Technology Research Center of Spectrum and Imaging Instrument, Electronic Information School, Wuhan University, Wuhan, 430072, P. R…. Hauptthemen laut MEDLINE: Ascorbic Acid, Point-of-Care Testing, Biomimetic Materials, Electrochemical Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Gefördert von: Key Technologies Research and Development Program.
Abstract (Originalsprache)
In the era of “Great Health”, maintaining an appropriate level of vitamin C is crucial for human health as it directly impacts the proper functioning of the immune system. In this study, a handheld nonenzymatic electrochemical sensor based on biomimetic material prussian blue functionalized polydopamine is proposed for point‐of‐care testing (POCT) of vitamin C in human fluids. Particularly, Prussian blue, known as a bio‐antidote, and polydopamine, a main component of cuttlefish ink, ensure the safety and reliability of home testing. The experimental results quantitatively demonstrate that this electrochemical POCT sensor enables rapid and accurate determination of vitamin C with a wide linear detection range from 0.2 to 200 µ m , a low limit of detection at 0.03 µ m , and high sensitivity at 0.33 µA µ m −1 cm −2 . More importantly, the performance evaluation using human urine and saliva samples confirms satisfactory accuracy and recovery rates for vitamin C determination by this electrochemical POCT sensor. Thus, the proposed handheld electrochemical POCT sensor holds potential utility as an affordable tool for VC determination and home‐based healthcare.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Hubei Engineering Technology Research Center of Spectrum and Imaging Instrument, Electronic Information School, Wuhan University, Wuhan, 430072, P. R. China.
- Department of Nephrology, Zhongnan Hospital of Wuhan University, Wuhan University, Wuhan, 430072, P. R. China.
- Wuhan Institute of Quantum Technology, Wuhan, 430206, P. R. China.
- Wuhan Danyaxiang Biological Technology Co. Ltd., Wuhan, 432000, P. R. China.
- College of Chemistry and Material Science, Shandong Agricultural University, Taian, Shandong, 271018, P. R. China.
ORCID-Profile: Xinghai Liu
Förderung
- Key Technologies Research and Development Program
Zitieren
APA
Fu, D., Dong, Y., Yin, J., Zhang, B., Zhang, S., Deng, J., Shui, H., & Liu, X. (2025). A Multi‐Channel Handheld Diagnostic Device Based on Green Biomimetic Material for Point‐of‐Care Testing of Vitamin C in Human Fluids. Advanced Healthcare Materials. https://doi.org/10.1002/adhm.202403454
Vancouver
Fu D, Dong Y, Yin J, Zhang B, Zhang S, Deng J, et al. A Multi‐Channel Handheld Diagnostic Device Based on Green Biomimetic Material for Point‐of‐Care Testing of Vitamin C in Human Fluids. Advanced Healthcare Materials. 2025. doi:10.1002/adhm.202403454
BibTeX
@article{Fu2025,
author = {Fu, Donglei and Dong, Yueyan and Yin, Junping and Zhang, Bowen and Zhang, Shuaibo and Deng, Junjie and Shui, Hua and Liu, Xinghai},
title = {A Multi‐Channel Handheld Diagnostic Device Based on Green Biomimetic Material for Point‐of‐Care Testing of Vitamin C in Human Fluids},
journal = {Advanced Healthcare Materials},
year = {2025},
doi = {10.1002/adhm.202403454}
}
EUVIMED