Accuracy of artificial intelligence software for CT angiography in stroke
Annals of Clinical and Translational Neurology · 2023
DOI: 10.1002/acn3.51790 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Clinical and Translational Neurology, 2023. Verfasst von Grant Mair und 11 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Clinical Brain Sciences, University of Edinburgh, Edinburgh, UK. Hauptthemen laut MEDLINE: Stroke, Arterial Occlusive Diseases. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical Research Council, Stroke Association, UK Dementia Research Institute.
Abstract (Originalsprache)
Objective
Software developed using artificial intelligence may automatically identify arterial occlusion and provide collateral vessel scoring on CT angiography (CTA) performed acutely for ischemic stroke. We aimed to assess the diagnostic accuracy of e‐CTA by Brainomix™ Ltd by large‐scale independent testing using expert reading as the reference standard.
Methods
We identified a large clinically representative sample of baseline CTA from 6 studies that recruited patients with acute stroke symptoms involving any arterial territory. We compared e‐CTA results with masked expert interpretation of the same scans for the presence and location of laterality‐matched arterial occlusion and/or abnormal collateral score combined into a single measure of arterial abnormality. We tested the diagnostic accuracy of e‐CTA for identifying any arterial abnormality (and in a sensitivity analysis compliant with the manufacturer's guidance that software only be used to assess the anterior circulation).
Results
We include CTA from 668 patients (50% female; median: age 71 years, NIHSS 9, 2.3 h from stroke onset). Experts identified arterial occlusion in 365 patients (55%); most (343, 94%) involved the anterior circulation. Software successfully processed 545/668 (82%) CTAs. The sensitivity, specificity and diagnostic accuracy of e‐CTA for detecting arterial abnormality were each 72% (95% CI = 66–77%). Diagnostic accuracy was non‐significantly improved in a sensitivity analysis excluding occlusions from outside the anterior circulation (76%, 95% CI = 72–80%) .
Interpretation
Compared to experts, the diagnostic accuracy of e‐CTA for identifying acute arterial abnormality was 72–76%. Users of e‐CTA should be competent in CTA interpretation to ensure all potential thrombectomy candidates are identified.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Centre for Clinical Brain Sciences, University of Edinburgh, Edinburgh, UK.
- Translational and Clinical Research Institute, Newcastle University, Newcastle Upon Tyne Hospitals NHS Trust, Newcastle upon Tyne, UK.
- Stroke Trials Unit, Mental Health & Clinical Neuroscience, University of Nottingham, Nottingham, UK.
- Institute of Neuroscience & Psychology, University of Glasgow, Glasgow, UK.
- Department of Neuroradiology, University Hospital, Technische Universität Dresden, Dresden, Germany.
- UK Dementia Research Institute Centre at the University of Edinburgh, Edinburgh, UK.
ORCID-Profile: Grant Mair, Philip White, Philip M Bath, Keith Muir, David Dye, Francesca Chappell, Rüdiger von Kummer, Malcolm Macleod, Nikola Sprigg, Joanna M Wardlaw
Förderung
- Medical Research Council
- Stroke Association
- UK Dementia Research Institute
- Alzheimer's Society
Zitieren
APA
Mair, G., White, P., Bath, P. M., Muir, K., Martin, C., Dye, D., Chappell, F., von Kummer, R., Macleod, M., Sprigg, N., Wardlaw, J. M., & RITeS Collaboration (2023). Accuracy of artificial intelligence software for CT angiography in stroke. Annals of Clinical and Translational Neurology. https://doi.org/10.1002/acn3.51790
Vancouver
Mair G, White P, Bath PM, Muir K, Martin C, Dye D, et al. Accuracy of artificial intelligence software for CT angiography in stroke. Annals of Clinical and Translational Neurology. 2023. doi:10.1002/acn3.51790
BibTeX
@article{Mair2023,
author = {Mair, Grant and White, Philip and Bath, Philip M and Muir, Keith and Martin, Chloe and Dye, David and Chappell, Francesca and von Kummer, Rüdiger and Macleod, Malcolm and Sprigg, Nikola and Wardlaw, Joanna M and RITeS Collaboration, RITeS Collaboration},
title = {Accuracy of artificial intelligence software for CT angiography in stroke},
journal = {Annals of Clinical and Translational Neurology},
year = {2023},
doi = {10.1002/acn3.51790}
}