Dose‑dependent image quality behavior of recently introduced clinical deep learning CT reconstruction methods

Joseph G. Meier, Elanchezhian Somasundaram, Samuel L. Brady

Journal of Applied Clinical Medical Physics · 2026-12 · 27 (12)

DOI: 10.1002/acm2.70917 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 19.11.2026. Verfasst von Joseph G. Meier und 2 weiteren Autor:innen. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Deep learning reconstruction (DLR) algorithms are increasingly used in CT to reduce image noise and support radiation dose reduction. However, DLR methods may alter spatial resolution and noise texture depending on dose level, reconstruction kernel, matrix size, and vendor implementation.

Purpose

To characterize how decreasing radiation dose affects noise magnitude, noise texture, and task‐based spatial resolution for two recently available clinical CT deep learning reconstruction algorithms: Canon Precise IQ Engine (PIQE) super‐resolution DLR and GE HealthCare TrueFidelity DLR.

Methods

The American College of Radiology CT accreditation phantom was scanned on an Aquilion ONE/INSIGHT Edition scanner (Canon Medical Systems USA) and a Revolution Apex scanner (GE HealthCare) across 12 dose levels. CTDIvol ranged from 0.7 to 20.0 mGy for Canon and from 0.68 to 20.46 mGy for GE. Canon images were reconstructed using filtered back projection (FBP) with a 512 matrix and PIQE Levels 1–3 with Body, Lung, and Cardiac kernels using a 1024 matrix. GE images were reconstructed using FBP and TrueFidelity low, medium, and high strengths with Standard, Lung, Bone, and Bone Plus kernels using a 512 matrix. Task‐based spatial resolution was evaluated using the task transfer function (TTF). Noise magnitude and texture were evaluated using the noise power spectrum (NPS), including the Noise Magnitude Ratio (NMR), defined as the square root of the ratio of NPS areas between the DLR and the FBP reference, and the Central Frequency Ratio (CFR), defined as the ratio of the center of gravity of the NPS between DLR and FBP, reflecting noise texture shifts.

Results

At a high‐dose comparison of 16.0 mGy for Canon and 16.43 mGy for GE, PIQE produced kernel‐dependent changes in bone TTF50% relative to matched FBP, ranging from a 21.3% decrease to a 83.6% increase across kernels and DLR strengths. TrueFidelity increased bone TTF50% relative to matched FBP moderately across all evaluated kernels and strengths, with changes ranging from 3.7% to 14.7%. As dose decreased, PIQE TTF50% and TTF10% decreased substantially across evaluated kernels and inserts, whereas TrueFidelity TTF values remained relatively stable. PIQE produced lower NMR values than TrueFidelity. With decreasing dose, PIQE also demonstrated decreasing NMR and CFR values, indicating greater relative noise reduction accompanied by a shift toward lower‐frequency noise texture.

Conclusions

PIQE and TrueFidelity demonstrated distinct dose‐dependent image quality behavior. PIQE provided substantial noise reduction and high task‐based spatial resolution at higher dose levels, consistent with its 1024 matrix super‐resolution implementation, but showed stronger dose dependence in TTF, NMR, and CFR. TrueFidelity demonstrated more stable spatial resolution and noise texture across dose levels, with more modest noise reduction. Although newer DLR algorithms demonstrate substantial potential for noise reduction, their effects on spatial resolution, noise magnitude, and noise texture must be characterized before aggressive dose reductions are implemented.

Zitieren

APA

Meier, J. G., Somasundaram, E., & Brady, S. L. (2026). Dose‑dependent image quality behavior of recently introduced clinical deep learning CT reconstruction methods. Journal of Applied Clinical Medical Physics, 27(12). https://doi.org/10.1002/acm2.70917

Vancouver

Meier JG, Somasundaram E, Brady SL. Dose‑dependent image quality behavior of recently introduced clinical deep learning CT reconstruction methods. Journal of Applied Clinical Medical Physics. 2026;27(12). doi:10.1002/acm2.70917

BibTeX
@article{Meier2026,
  author = {Meier, Joseph G. and Somasundaram, Elanchezhian and Brady, Samuel L.},
  title = {Dose‑dependent image quality behavior of recently introduced clinical deep learning CT reconstruction methods},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {12},
  doi = {10.1002/acm2.70917}
}
Kostenlos starten →