Accuracy of reference air kerma in estimating incident air kerma at skin in interventional neuroradiology: Effect of X‐ray field size and alternative PERP corrections

Atsushi Fukuda, Nao Ichikawa, Takuma Hayashi, Pei-Jan Paul Lin iD, Ayaka Hirosawa iD, Kosuke Matsubara iD

Journal of Applied Clinical Medical Physics · 2026-11 · 27 (11)

DOI: 10.1002/acm2.70903 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 11.11.2026. Verfasst von Atsushi Fukuda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Fukushima, Japan. Hauptthemen laut MEDLINE: Phantoms, Imaging, Skin, Radiology, Interventional, Radiometry. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Kerma‐area product (KAP) meters monitor reference air kerma () and KAP (. While is calculated using and collimator positions, is directly measured via a small central ionization chamber. Although is a common surrogate for incident air kerma at the skin (), its clinical application to interventional neuroradiological procedures carries higher uncertainty than regular interventional radiology, with source‐to‐skin distance being the greatest factor due to the smaller head size. Additionally, may not perfectly represent actual due to variations in beam quality, filtration, and X‐ray field size.

Purpose

This study investigated (1) the accuracy of and , and (2) the feasibility of employing them as surrogates for during cerebral angiography and interventional neuroradiological procedures.

Methods

A solid‐state dosimeter (AGMS‐DM+) was placed at the patient entrance reference point (PERP) on a polymethyl methacrylate phantom to measure . Accuracies of and were evaluated against in two investigations. First, effects of tube voltage and spectral shaping filters (1.2 mmAl, 0.03 mmTa, and 0.2 mmCu) were evaluated with/without equalization filters inserted symmetrically by 2.5 cm from both sides at a fixed X‐ray field size of 10 × 10 cm 2 . Second, the effect of varying the X‐ray field size was investigated for each filter. Additionally, a fluoroscopic system and a head phantom were modeled using computer‐aided design software. Three fields of view (FOVs) were evaluated to simulate different clinical stages: 30 × 30 cm 2 (for whole angiography), 15 × 15 cm 2 , and 3.8 × 3.8 cm 2 (for targeted interventional neuroradiology). Target centers were aligned with the brain center and precallosal segment of the anterior cerebral artery. Six typical clinical projections were used per FOV to determine source‐to‐skin distance and X‐ray field size at the PERP, allowing calculation of the skin‐level errors for both and . The effects of the corrections were also investigated using the X‐ray field size measured by / and an alternative PERP (aPERP) positioned 11 cm from the isocenter toward the X‐ray focus.

Results

Relative errors for both methods as functions of tube voltage across the three filters were within ± 4.0%. With equalization filters, errors ranged from −49.0%–−33.9%, exceeding the acceptable accuracy tolerance (± 35%), whereas errors (−15.8%–−10.8%) remained within the tolerance. The relative errors of and as a function of the X‐ray field size ranged from −9.7%–12.7% and −58.3%–26.1%, respectively. Although most skin‐level errors for both and remained within the ± 35% tolerance in clinical simulation, those for at the 3.8 × 3.8 cm 2 FOV exceeded it. However, with the X‐ray field size and aPERP corrections, the errors for were reduced to within ± 2.7%. Conclusions and should be strategically selected based on the specific clinical application, such as standard cerebral angiography versus targeted interventional neuroradiology. Particularly, provides superior accuracy when equalization filters are used or when appropriate corrections for X‐ray field size and aPERP are applied in small‐field dosimetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingSkinRadiology, InterventionalRadiometryRadiography, InterventionalHumansRadiation DosageX-RaysImage Processing, Computer-Assisted

Stichwörter der Autor:innen

incident air kerma at skininterventional neuroradiologykerma‐area productreference air kerma

Autor:innen und Institutionen

  • Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Fukushima, Japan.
  • Department of Radiological Technology, Faculty of Health Science, Kobe Tokiwa University, Kobe, Hyogo, Japan.
  • Department of Radiology, Shiga Psychiatric Medical Center, Kusatsu, Shiga, Japan.
  • Department of Radiology, Virginia Commonwealth University Health System, Richmond, Virginia, USA.
  • Department of Medical Technology, Toyama Prefectural Central Hospital, Toyama, Toyama, Japan.
  • Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.

ORCID-Profile: Pei-Jan Paul Lin, Ayaka Hirosawa, Kosuke Matsubara

Zitieren

APA

Fukuda, A., Ichikawa, N., Hayashi, T., Lin, P. P., Hirosawa, A., & Matsubara, K. (2026). Accuracy of reference air kerma in estimating incident air kerma at skin in interventional neuroradiology: Effect of X‐ray field size and alternative PERP corrections. Journal of Applied Clinical Medical Physics, 27(11). https://doi.org/10.1002/acm2.70903

Vancouver

Fukuda A, Ichikawa N, Hayashi T, Lin PP, Hirosawa A, Matsubara K. Accuracy of reference air kerma in estimating incident air kerma at skin in interventional neuroradiology: Effect of X‐ray field size and alternative PERP corrections. Journal of Applied Clinical Medical Physics. 2026;27(11). doi:10.1002/acm2.70903

BibTeX
@article{Fukuda2026,
  author = {Fukuda, Atsushi and Ichikawa, Nao and Hayashi, Takuma and Lin, Pei-Jan Paul and Hirosawa, Ayaka and Matsubara, Kosuke},
  title = {Accuracy of reference air kerma in estimating incident air kerma at skin in interventional neuroradiology: Effect of X‐ray field size and alternative PERP corrections},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {11},
  doi = {10.1002/acm2.70903}
}
Kostenlos starten →