Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance

Samira Aghaei Akizi iD, Seyed Mahmoud Reza Aghamiri, Zahra Siavashpour iD, Reza Ghaderi, Soharb Barimani

Journal of Applied Clinical Medical Physics · 2026-11 · 27 (11)

DOI: 10.1002/acm2.70882 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 27.10.2026. Verfasst von Samira Aghaei Akizi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Radiation Engineering, Shahid Beheshti University, Tehran, Iran. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Quality Assurance, Health Care, Machine Learning. Die Arbeit nennt 72 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accurate positioning of multi‐leaf collimators (MLCs) is critical for ensuring the dosimetric precision of intensity‐modulated radiation therapy (IMRT). Nonetheless, daily MLC positional variations can degrade treatment accuracy, emphasizing the need for robust methodologies to detect and predict these deviations.

Purpose

This study proposes a machine learning (ML) approach to predict daily MLC positional deviations using factors contributing to positional errors, enabling daily patient‐specific quality assurance (PSQA) to assess the dosimetric impact of these discrepancies on patient treatment.

Methods

DICOM‐RT and Dynalog files from 29 IMRT patients were collected, and 78 features were extracted, including aperture‐based features, mechanical parameters, and MLC check data. A LightGBM machine learning model was constructed to predict daily positional errors between planned and delivered MLC positions at each control point. For dosimetric analysis, nine independent IMRT plans were evaluated using dose volume histogram (DVH) parameters such as D 98% , and V 100% for targets and D max , D mean for Organs at Risk (OARs). Moreover, the Gamma Pass Rate (GPR) was calculated for both the delivered and predicted dose distributions, with regarding the primary planned dose distribution as the reference.

Results

The proposed LightGBM model demonstrated accurate predictions, achieving Mean Absolute Error (MAE) and Root Mean Square Error (RMSE) of 0.035 mm and 0.078 mm for the training dataset, 0.070 mm and 0.164 mm for the validation dataset, and 0.054 mm and 0.144 mm for the test dataset, respectively. Leaf velocity, with a Spearman correlation coefficient of 0.703, was the most influential feature. Evaluation on nine independent clinical cases demonstrated good agreement between delivered and predicted dosimetric deviations from the planned values, with mean differences below 0.200 Gy and approximately 1.5% for target structures, and even smaller discrepancies for OARs. The GPR (3%/3 mm criterion) was about 100% for delivered doses and 99.99% for predicted doses.

Conclusions

The proposed ML‐based LightGBM model offers an efficient daily PSQA tool, with the potential to enhance the treatment accuracy and support proactive clinical decision‐making in IMRT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareMachine LearningOrgans at RiskParticle AcceleratorsNeoplasmsHumansRadiotherapy DosageBoosting Machine Learning AlgorithmsPredictive Learning Models

Stichwörter der Autor:innen

daily MLC positional errorsdaily dosimetric analysis in IMRTmachine learning

Autor:innen und Institutionen

  • Department of Medical Radiation Engineering, Shahid Beheshti University, Tehran, Iran.
  • Department of Radiation Oncology, Cancer Institute, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Clinical Research Development Unit, Shohada-e Tajrish Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Electrical Engineering, Shahid Beheshti University, Tehran, Iran.
  • Department of Biomedical Engineering, Amirkabir University of Technology, Tehran, Iran.

ORCID-Profile: Samira Aghaei Akizi, Zahra Siavashpour

Zitieren

APA

Akizi, S. A., Aghamiri, S. M. R., Siavashpour, Z., Ghaderi, R., & Barimani, S. (2026). Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance. Journal of Applied Clinical Medical Physics, 27(11). https://doi.org/10.1002/acm2.70882

Vancouver

Akizi SA, Aghamiri SMR, Siavashpour Z, Ghaderi R, Barimani S. Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance. Journal of Applied Clinical Medical Physics. 2026;27(11). doi:10.1002/acm2.70882

BibTeX
@article{Akizi2026,
  author = {Akizi, Samira Aghaei and Aghamiri, Seyed Mahmoud Reza and Siavashpour, Zahra and Ghaderi, Reza and Barimani, Soharb},
  title = {Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {11},
  doi = {10.1002/acm2.70882}
}
Kostenlos starten →