Quantifying setup uncertainty between computed tomography guidance and magnetic resonance guidance in intramuscular metastases radiotherapy
Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)
DOI: 10.1002/acm2.70589 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 19.04.2026. Verfasst von Travis Salzillo und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Muscle Neoplasms, Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging, Radiotherapy, Image-Guided. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, National Institutes of Health/National Cancer Institute.
Abstract (Originalsprache)
Background and purpose MR‐Linac (MRL) is well‐suited for treatment of oligometastatic disease (OMD) in muscle due to its superior soft tissue contrast. The purpose of this study was to assess dosimetric differences and quantify margins when CT‐guidance (CTgRT) is used for OMD treatment compared to MR‐guidance (MRgRT) on the MRL. Materials and methods Five patients with intramuscular oligometastasis were treated with the MRL using daily MRgRT. To simulate CT‐based alignment, an in‐house deep learning model was used to generate synthetic‐CT images from MR images for each fraction. The synthetic‐CTs were independently aligned to the corresponding simulation‐CTs by seven physicists using rigid registration. The difference between CTgRT and MRgRT alignment was calculated to quantify the inter‐fractional image‐guidance uncertainty and estimate PTV margins. Dose from the clinical beam sets was recalculated on the synthetic‐CTs for comparative analysis.
Results
The CTgRT alignment compared to MRgRT were −0.59 ± 3.49 mm, 2.04 ± 3.96 mm, and 1.23 ± 1.20 mm in left‐right (LR), superior‐inferior (SI), and anterior‐posterior (AP) directions, respectively. Compared to clinical plans, the dosimetric parameters of recalculated synthetic‐CT plans, V95 and D95, were significantly lower, while the homogeneity index was significantly higher ( p < 0.001 for each metric). Compared to MRgRT, CTgRT required additional treatment margins of 10.0 mm in LR, 11.5 mm in SI, and 4.6 mm in AP directions.
Conclusions
We quantified the dosimetric difference in treating intramuscular metastases using MRgRT versus CTgRT and showed that treatment margin can be reduced by at least 5 mm in each direction with MRgRT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Genitourinary Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
ORCID-Profile: Travis Salzillo, Surendra Prajapati, Zhiqian Henry Yu, He Wang, Xinru Chen, Jinzhong Yang
Förderung
- NCI NIH HHS
- National Institutes of Health/National Cancer Institute
Zitieren
APA
Salzillo, T., Prajapati, S., Yu, Z. H., Zhao, Y., Lim, T. Y., Wang, H., Wang, X. A., Chen, X., Kudchadker, R., Lee, B., Sobremonte, A., Farooqi, A. S., Mitra, D., Bishop, A., Tang, C., & Yang, J. (2026). Quantifying setup uncertainty between computed tomography guidance and magnetic resonance guidance in intramuscular metastases radiotherapy. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70589
Vancouver
Salzillo T, Prajapati S, Yu ZH, Zhao Y, Lim TY, Wang H, et al. Quantifying setup uncertainty between computed tomography guidance and magnetic resonance guidance in intramuscular metastases radiotherapy. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70589
BibTeX
@article{Salzillo2026,
author = {Salzillo, Travis and Prajapati, Surendra and Yu, Zhiqian Henry and Zhao, Yao and Lim, Tze Yee and Wang, He and Wang, Xin A and Chen, Xinru and Kudchadker, Rajat and Lee, Belinda and Sobremonte, Angela and Farooqi, Ahsan S and Mitra, Devarati and Bishop, Andrew and Tang, Chad and Yang, Jinzhong},
title = {Quantifying setup uncertainty between computed tomography guidance and magnetic resonance guidance in intramuscular metastases radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {4},
doi = {10.1002/acm2.70589}
}
EUVIMED