Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Marte Kåstad Høiskar und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Norwegian University of Science and Technology, Trondheim, Norway. Hauptthemen laut MEDLINE: Head and Neck Neoplasms, Proton Therapy, Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accurate target delineation is essential for precise delivery of proton therapy. A magnetic resonance imaging (MRI)‐only radiotherapy workflow may improve delineations, and thus improve proton therapy, but requires reliable synthetic computed tomography (sCT) for accurate dose calculations.

Purpose

The aim was to evaluate the dosimetric accuracy of commercial software‐based sCT for proton dose calculation for head and neck cancer (HNC), and benchmark it against the accuracy of photon dose calculation on the same sCT.

Methods

MRI and planning CT (pCT) were acquired for 20 HNC patients before receiving photon therapy. A commercial software created MRI‐based sCTs, which were co‐registered to pCTs. For each patient, photon and proton plans based on pCT, prescribing 68 Gy to the target, were created in RayStation and recalculated on sCT. Dose‐volume histogram (DVH) metrics and gamma index (2%/2 mm criteria) were analyzed, comparing pCT‐ and sCT‐calculated dose. Local and global gamma index were calculated for low and high‐dose thresholds. A gamma pass rate (GPR) was calculated for each plan and gamma index type.

Results

For photon and proton plans, the median difference in DVH metrics between pCT‐ and sCT‐calculated dose was < 0.4 Gy with an interquartile range < ± 0.7 Gy for all structures and metrics, except for mean dose to oesophagus (median = −0.3 Gy, range: −2.0 Gy to 0.7 Gy), oral cavity (median = 1.8 Gy, range: 0.6 Gy to 3.0 Gy), and larynx (median = −0.1 Gy, range: −1.5 Gy to 0.6 Gy) for proton plans. The median GPR for photon plans was > 97.7%, while for proton plans it was > 93.3% except for body GPR with low‐dose threshold.

Conclusion

While sCT seems feasible for photon therapy, poorer agreement between pCT‐ and sCT‐calculated proton dose was found. Patient positioning differences between CT and MRI may partly explain the discrepancies between pCT‐ and sCT‐calculated proton dose.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Head and Neck NeoplasmsProton TherapyRadiotherapy Planning, Computer-AssistedMagnetic Resonance ImagingTomography, X-Ray ComputedPhotonsHumansRadiotherapy DosageOrgans at RiskRadiotherapy, Intensity-ModulatedRadiometryImage Processing, Computer-Assisted

Stichwörter der Autor:innen

Head and neck cancerMRI‐based synthetic CTProton therapy

Autor:innen und Institutionen

  • Department of Physics, Norwegian University of Science and Technology, Trondheim, Norway.
  • Section for Biomedical Physics, Department of Radiation Oncology, University Hospital and Medical Faculty, Eberhard Karls University Tübingen, Geschwister-Scholl-Platz, Tübingen, Germany.
  • Department of Radiotherapy, Cancer Clinic, St. Olavs Hospital, Trondheim University Hospital, Trondheim, Norway.

Zitieren

APA

Høiskar, M. K., Winter, R. M., Hornik, N., Alsaker, M. D., Fridstrøm, K., Almberg, S. S., & Redalen, K. R. (2025). Dosimetric evaluation of magnetic resonance imaging based synthetic computed tomography for head and neck photon and proton therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70335

Vancouver

Høiskar MK, Winter RM, Hornik N, Alsaker MD, Fridstrøm K, Almberg SS, et al. Dosimetric evaluation of magnetic resonance imaging based synthetic computed tomography for head and neck photon and proton therapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70335

BibTeX
@article{Hiskar2025,
  author = {Høiskar, Marte Kåstad and Winter, René M and Hornik, Natalie and Alsaker, Mirjam Delange and Fridstrøm, Kajsa and Almberg, Sigrun Saur and Redalen, Kathrine Røe},
  title = {Dosimetric evaluation of magnetic resonance imaging based synthetic computed tomography for head and neck photon and proton therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70335}
}
Kostenlos starten →