Enhancing patient‐specific quality assurance in MR‐guided radiation therapy: A fluence‐based method using log files
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70238 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Kai Yuan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Division, Department of Medical Innovation & Technology, CUHK Medical Centre, Hong Kong SAR, China. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Patient‐specific quality assurance (PSQA) is crucial in radiation therapy to ensure accurate and safe dose delivery. The Elekta Unity MR‐Linac system, which combines MRI with a linear accelerator, presents unique challenges for conventional PSQA methods due to its adaptive capabilities and the presence of a magnetic field.
Purpose
This study introduced a novel PSQA method for the Elekta Unity MR‐Linac system, utilizing treatment log files and fluence map verification to provide a more efficient alternative to traditional measurement‐based techniques.
Methods
The proposed method analyzed log file data, including monitor units (MU), multi‐leaf collimator (MLC) positions, and jaw positions, comparing them with treatment planning system (TPS) values. Fluence maps were generated for each gantry angle and evaluated using gamma analysis. A user‐friendly interface was developed to streamline the process.
Results
The method was tested on nine stereotactic body radiation therapy (SBRT) cases, showing strong concordance between planned and delivered parameters. MU deviations were within 0.5 MU, X‐jaw and MLC leaf position deviations were under 1 mm, and gamma analysis of fluence maps achieved passing rates above 98% for all gantry angles.
Conclusion
This log file‐based PSQA method offered distinct advantages over traditional measurement‐based approaches, including reduced QA time, direct assessment of the delivered plan, and comprehensive evaluation of treatment parameters. This method provided an efficient and accurate alternative PSQA solution for MR‐guided adaptive radiotherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics Division, Department of Medical Innovation & Technology, CUHK Medical Centre, Hong Kong SAR, China.
- Department of Clinical Oncology, Faculty of Medicine, The Chinese University of Hong Kong, Hong Kong SAR, China.
ORCID-Profile: Kai Yuan
Zitieren
APA
Yuan, K., Man-Hin Cheung, M., & Lee, L. (2025). Enhancing patient‐specific quality assurance in MR‐guided radiation therapy: A fluence‐based method using log files. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70238
Vancouver
Yuan K, Man-Hin Cheung M, Lee L. Enhancing patient‐specific quality assurance in MR‐guided radiation therapy: A fluence‐based method using log files. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70238
BibTeX
@article{Yuan2025,
author = {Yuan, Kai and Man-Hin Cheung, Matthew and Lee, Louis},
title = {Enhancing patient‐specific quality assurance in MR‐guided radiation therapy: A fluence‐based method using log files},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70238}
}
EUVIMED