The impact of NaI(Tl) crystal hydration on gamma camera spectral response and image uniformity

Grace Eliason, Michael S Silosky iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70218 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Grace Eliason und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of Colorado School of Medicine, Aurora, Colarado, USA. Hauptthemen laut MEDLINE: Gamma Cameras, Sodium Iodide, Thallium, Water. Die Arbeit nennt 7 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

NaI(Tl) scintillators used in most gamma cameras are hermetically sealed to prevent the absorption of water molecules from the surrounding environment. If this seal is compromised, crystal hydration, a localized defect resulting in non‐uniform attenuation of scintillation photons, may occur.

Purpose

The purpose of this study was to evaluate the effects of crystal hydration across multiple radionuclides, characterizing the impact on spectral response and image uniformity.

Methods

Hydration was assessed using on‐peak and off‐peak imaging for several radionuclides. The progression of hydration was assessed by calculating the contrast‐to‐noise ratio in hydrated regions on monthly uniformity maps. Spectra were captured for both hydrated and non‐hydrated regions of the detector to determine photopeak energy and energy resolution.

Results

The visual appearance of the hydration effect in off‐peak images was greatest for 133 Xe. The effect was more substantial in the 10% low off‐peak image than in the 10% high off‐peak image. Hydration was not observed in on‐peak images for either 133 Xe or 99m Tc and was only barely visible for 131 I. CNR measurements show a slow but notable progression of hydration in uniformity maps over time. The measured photopeak in hydrated regions was lower than that of non‐hydrated regions. Hydration also resulted in a degradation in energy resolution with the effect being more significant at lower energies.

Conclusions

Hydrated regions of the detector demonstrated an energy‐dependent degradation in energy resolution which corresponded to an energy dependence in the detectability of hydration in off‐peak images. Monthly updates to the uniformity correction maps were able to adequately account for hydration. When hydration has been identified, routine evaluation of its impact on uniformity maps may provide a convenient method of tracking progression.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gamma CamerasSodium IodideThalliumWaterImage Processing, Computer-AssistedPhantoms, ImagingRadionuclide ImagingHumansIodidesSodium

Stichwörter der Autor:innen

NaI(Tl)QCgamma camerahydrationnuclear medicine

Substanzen

Sodium IodideThalliumWaterthallium-doped sodium iodideIodidesSodium

Autor:innen und Institutionen

  • Department of Radiology, University of Colorado School of Medicine, Aurora, Colarado, USA.

ORCID-Profile: Michael S Silosky

Zitieren

APA

Eliason, G., & Silosky, M. S. (2025). The impact of NaI(Tl) crystal hydration on gamma camera spectral response and image uniformity. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70218

Vancouver

Eliason G, Silosky MS. The impact of NaI(Tl) crystal hydration on gamma camera spectral response and image uniformity. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70218

BibTeX
@article{Eliason2025,
  author = {Eliason, Grace and Silosky, Michael S},
  title = {The impact of NaI(Tl) crystal hydration on gamma camera spectral response and image uniformity},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70218}
}
Kostenlos starten →