Assessment of the impact of CT scan protocols on carbon ion radiotherapy treatment planning
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70203 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Yuya Miyasaka und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Tomography, X-Ray Computed, Heavy Ion Radiotherapy, Head and Neck Neoplasms. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.
Abstract (Originalsprache)
Purpose
The purpose of this study was to quantify the impact of computed tomography (CT) scan protocols and CT scanner on carbon ion radiotherapy treatment planning Material and Methods A table was created for each of the five parameters to relate CT values to stopping power (SPR) (CT‐SPR conversion table). The parameters changed in the four protocols, with the exception of the clinical protocol, are as follows: tube voltage, convolution kernel, CT scanner, and CT model. Treatment plans were calculated with each of the five CT‐SPR conversion tables applied, and differences from the doses in the clinical protocol were evaluated by comparison of DVH parameters and gamma analysis.
Results
The largest difference in CT values in the protocols with changing tube voltage compared to the clinical protocol was more than 200 HU for the high CT value material. The change in CT values due to the change in the reconstructed kernel was slight, and the difference in CT values for the CT scanner change or model change protocol was similar. Mean difference
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan.
- Department of Radiation Oncology, Yamagata University Faculty of Medicine, Yamagata, Japan.
- Department of Radiology, Nihon-kai General Hospital, Yamagata, Japan.
ORCID-Profile: Yasuhito Hagiwara, Miyu Ishizawa, Hiraku Sato
Förderung
- JSPS KAKENHI
Zitieren
APA
Miyasaka, Y., Souda, H., Hagiwara, Y., Akamatsu, H., Harada, M., Kaneko, T., Chai, H., Ishizawa, M., Sato, H., Koto, M., & Iwai, T. (2025). Assessment of the impact of CT scan protocols on carbon ion radiotherapy treatment planning. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70203
Vancouver
Miyasaka Y, Souda H, Hagiwara Y, Akamatsu H, Harada M, Kaneko T, et al. Assessment of the impact of CT scan protocols on carbon ion radiotherapy treatment planning. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70203
BibTeX
@article{Miyasaka2025,
author = {Miyasaka, Yuya and Souda, Hikaru and Hagiwara, Yasuhito and Akamatsu, Hiroko and Harada, Mayumi and Kaneko, Takashi and Chai, Hongbo and Ishizawa, Miyu and Sato, Hiraku and Koto, Masashi and Iwai, Takeo},
title = {Assessment of the impact of CT scan protocols on carbon ion radiotherapy treatment planning},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70203}
}
EUVIMED