EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Quality assurance of craniospinal irradiation in helical tomotherapy using a dose reconstruction tool based on leaf open time

Xiunan Wang iD, Wenjie Ni, Yongqing Ge, Bo Liu, Qin Wang, Linan Song, Hui Yang, Yizhen Jin, Xiaofeng Mu iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70139 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Xiunan Wang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Beijing Shijitan Hospital of Capital Medical University, Beijing, China. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care, Craniospinal Irradiation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The delivery quality assurance (DQA) of craniospinal irradiation (CSI) due to the target length results in no ideal verification devices. Delivery Analysis (DA) could calculate the dose distribution based on the measured multi leaf open time in helical tomotherapy (HT). This study aimed to evaluate the efficacy of DA for DQA of CSI in HT. Material and Methods 32 CSI plans were classified into two groups based on γ analysis of the PTV‐cranial and PTV‐spine plans using a 2D ionization chamber matrix (MatriXX). Plans with γ passing rates ≥ 95% at 3%/2 mm were classified as the passed group, while those < 95% were classified as the failed group. Receiver operating characteristic (ROC) curves identified optimal passing rate threshold for DA in HT. Logistic regression analyzed risk factors for DQA failure, and failed plans were reoptimized according to the adjusted parameter.

Results

For PTV‐cranial plans, 30 passed and two failed; for PTV‐spine plans, 21 passed and 11 failed. ROC analysis revealed areas under the curve of 0.858 (PTV‐cranial, threshold: 89.0%) and 0.714 (PTV‐spine, threshold: 86.0%). Logistic regression identified planned modulation factor (MF‐plan; p = 0.046; p = 0.023) and actual modulation factor (MF‐actual; p = 0.027; p = 0.008) as independent risk factors for DQA failure in both MatriXX and DA. Additionally, beam on time ( p = 0.043), gantry period ( p = 0.007) and maximum leaf open time ( p = 0.007) were identified as independent risk factors for DA. Reoptimization of failed plans with MF‐plan = 2.6 significantly improved passing rates in DA (73.70% ± 13.30% vs. 88.20% ± 12.30%; p = 0.010) and MatriXX (91.20% ± 2.60% vs. 96.10% ± 1.40%; p < 0.001).

Conclusion

Delivery Analysis could be a feasible tool for DQA of CSI in HT. Increasing the MF‐plan is recommended to enhance the passing rate.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedQuality Assurance, Health CareCraniospinal IrradiationHumansRadiotherapy DosageOrgans at RiskChildMaleFemaleChild, Preschool

Stichwörter der Autor:innen

craniospinal irradiationdose reconstructionhelical tomotherapyleaf open time

Autor:innen und Institutionen

  • Department of Radiation Oncology, Beijing Shijitan Hospital of Capital Medical University, Beijing, China.

ORCID-Profile: Xiunan Wang, Xiaofeng Mu

Zitieren

APA

Wang, X., Ni, W., Ge, Y., Liu, B., Wang, Q., Song, L., Yang, H., Jin, Y., & Mu, X. (2025). Quality assurance of craniospinal irradiation in helical tomotherapy using a dose reconstruction tool based on leaf open time. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70139

Vancouver

Wang X, Ni W, Ge Y, Liu B, Wang Q, Song L, et al. Quality assurance of craniospinal irradiation in helical tomotherapy using a dose reconstruction tool based on leaf open time. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70139

BibTeX
@article{Wang2025,
  author = {Wang, Xiunan and Ni, Wenjie and Ge, Yongqing and Liu, Bo and Wang, Qin and Song, Linan and Yang, Hui and Jin, Yizhen and Mu, Xiaofeng},
  title = {Quality assurance of craniospinal irradiation in helical tomotherapy using a dose reconstruction tool based on leaf open time},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70139}
}
Kostenlos starten →