Clinical commissioning and introduction of an in‐house artificial intelligence (AI) platform for automated head and neck intensity modulated radiation therapy (IMRT) treatment planning
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14558 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Xinyi Li und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, United States. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Head and Neck Neoplasms, Radiotherapy Dosage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Background and purpose To describe the clinical commissioning of an in‐house artificial intelligence (AI) treatment planning platform for head‐and‐neck (HN) Intensity Modulated Radiation Therapy (IMRT). Materials and methods The AI planning platform has three components: (1) a graphical user interface (GUI) is built within the framework of a commercial treatment planning system (TPS). The GUI allows AI models to run remotely on a designated workstation configured with GPU acceleration. (2) A template plan is automatically prepared involving both clinical and AI considerations, which include contour evaluation, isocenter placement, and beam/collimator jaw placement. (3) A well‐orchestrated suite of AI models predicts optimal fluence maps, which are imported into TPS for dose calculation followed by an optional automatic fine‐tuning. Six AI models provide flexible tradeoffs in parotid sparing and Planning Target Volume (PTV)‐organ‐at‐risk (OAR) preferences. Planners could examine the plan dose distribution and make further modifications as clinically needed. The performance of the AI plans was compared to the corresponding clinical plans.
Results
The average plan generation time including manual operations was 10–15 min per case, with each AI model prediction taking ∼1 s. The six AI plans form a wide range of tradeoff choices between left and right parotids and between PTV and OARs compared with corresponding clinical plans, which correctly reflected their tradeoff designs.
Conclusion
The in‐house AI IMRT treatment planning platform was developed and is available for clinical use at our institution. The process demonstrates outstanding performance and robustness of the AI platform and provides sufficient validation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, United States.
- Department of Information Systems, University of North Carolina at Charlotte, Charlotte, North Carolina, United States.
- Department of Head and Neck Surgery and Communication Sciences, Duke University Medical Center, Durham, North Carolina, United States.
Förderung
- NCI NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Li, X., Sheng, Y., Wu, Q. J., Ge, Y., Brizel, D. M., Mowery, Y. M., Yang, D., Yin, F. F., & Wu, Q. (2025). Clinical commissioning and introduction of an in‐house artificial intelligence (AI) platform for automated head and neck intensity modulated radiation therapy (IMRT) treatment planning. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14558
Vancouver
Li X, Sheng Y, Wu QJ, Ge Y, Brizel DM, Mowery YM, et al. Clinical commissioning and introduction of an in‐house artificial intelligence (AI) platform for automated head and neck intensity modulated radiation therapy (IMRT) treatment planning. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14558
BibTeX
@article{Li2025,
author = {Li, Xinyi and Sheng, Yang and Wu, Qingrong Jackie and Ge, Yaorong and Brizel, David M and Mowery, Yvonne M and Yang, Dongrong and Yin, Fang-Fang and Wu, Qiuwen},
title = {Clinical commissioning and introduction of an in‐house artificial intelligence (AI) platform for automated head and neck intensity modulated radiation therapy (IMRT) treatment planning},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14558}
}
EUVIMED