Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Mustafa Kadhim und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Radiation Physics, Lund University, Lund, Sweden. Hauptthemen laut MEDLINE: Breath Holding, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Onkologische Pflege
Abstract (Originalsprache)
Purpose
This study investigated the use of surface guided radiotherapy (SGRT) in combination with a tomotherapy treatment mode using discrete delivery angles for deep inspiration breath hold (DIBH) treatments of breast cancer (bc). We aimed to assess the feasibility and dosimetric advantages of this approach. Materials and methods We evaluated camera occlusion in the Radixact treatment system bore and the stability of DIBH signals during couch movement. The SGRT system's ability to maintain signal and surface image accuracy was analyzed at different depths within the bore. Dosimetric parameters were compared and measured for 20 left‐sided bc patients receiving TomoDirect (TD) tangential radiotherapy in both DIBH and free breathing (FB).
Results
The SGRT system maintained surface coverage and precise DIBH‐signal at depths up to 40 cm beyond the treatment center. Camera occlusion occurred in the clavicular and neck regions due to the patient's morphology and gantry geometry. Nonetheless, the system accurately detected respiratory motion for all measurements. The DIBH plans significantly ( p < 0.001) reduced mean heart and left anterior descending artery (LAD) radiation doses by up to 40%, with a 50% reduction in near‐maximum heart and LAD doses, respectively. No significant dosimetric differences between DIBH and FB were observed in other investigated parameters and volumes.
Conclusions
Camera occlusion and couch movement minimally impacted the real‐time surface image accuracy needed for DIBH treatments of bc. DIBH reduced heart and LAD radiation doses significantly compared to FB, indicating the feasibility and dosimetric benefits of combining these modalities.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Radiation Physics, Lund University, Lund, Sweden.
- Radiation Physics, Department of Hematology, Oncology, and Radiation Physics, Skåne University Hospital, Lund, Sweden.
ORCID-Profile: Mustafa Kadhim
Zitieren
APA
Kadhim, M., Haraldsson, A., Kügele, M., Enocson, H., Bäck, S., & Ceberg, S. (2024). Surface guided ring gantry radiotherapy in deep inspiration breath hold for breast cancer patients. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14463
Vancouver
Kadhim M, Haraldsson A, Kügele M, Enocson H, Bäck S, Ceberg S. Surface guided ring gantry radiotherapy in deep inspiration breath hold for breast cancer patients. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14463
BibTeX
@article{Kadhim2024,
author = {Kadhim, Mustafa and Haraldsson, André and Kügele, Malin and Enocson, Hedda and Bäck, Sven and Ceberg, Sofie},
title = {Surface guided ring gantry radiotherapy in deep inspiration breath hold for breast cancer patients},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14463}
}
EUVIMED