Geometric and dosimetric evaluation for breast and regional nodal auto‐segmentation structures
Journal of Applied Clinical Medical Physics · 2024-10 · 25 (10)
DOI: 10.1002/acm2.14461 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 02.08.2024. Verfasst von Tiffany Tsui und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Loyola University Chicago, Stritch School of Medicine, Maywood, Illinois, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Breast Neoplasms, Radiotherapy Dosage, Mastectomy. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The accuracy of artificial intelligence (AI) generated contours for intact‐breast and post‐mastectomy radiotherapy plans was evaluated. Geometric and dosimetric comparisons were performed between auto‐contours (ACs) and manual‐contours (MCs) produced by physicians for target structures. Breast and regional nodal structures were manually delineated on 66 breast cancer patients. ACs were retrospectively generated. The characteristics of the breast/post‐mastectomy chestwall (CW) and regional nodal structures (axillary [AxN], supraclavicular [SC], internal mammary [IM]) were geometrically evaluated by Dice similarity coefficient (DSC), mean surface distance, and Hausdorff Distance. The structures were also evaluated dosimetrically by superimposing the MC clinically delivered plans onto the ACs to assess the impact of utilizing ACs with target dose (Vx%) evaluation. Positive geometric correlations between volume and DSC for intact‐breast, AxN, and CW were observed. Little or anti correlations between volume and DSC for IM and SC were shown. For intact‐breast plans, insignificant dosimetric differences between ACs and MCs were observed for AxN V95% ( p = 0.17) and SC V95% ( p = 0.16), while IMN V90% ACs and MCs were significantly different. The average V95% for intact‐breast MCs (98.4%) and ACs (97.1%) were comparable but statistically different ( p = 0.02). For post‐mastectomy plans, AxN V95% ( p = 0.35) and SC V95% ( p = 0.08) were consistent between ACs and MCs, while IMN V90% was significantly different. Additionally, 94.1% of AC‐breasts met ΔV95% variation 0.7. However, only 62.5% AC‐CWs achieved the same metrics, despite AC‐CW V95% ( p = 0.43) being statistically insignificant. The AC intact‐breast structure was dosimetrically similar to MCs. The AC AxN and SC may require manual adjustments. Careful review should be performed for AC post‐mastectomy CW and IMN before treatment planning. The findings of this study may guide the clinical decision‐making process for the utilization of AI‐driven ACs for intact‐breast and post‐mastectomy plans. Before clinical implementation of this auto‐segmentation software, an in‐depth assessment of agreement with each local facilities MCs is needed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Loyola University Chicago, Stritch School of Medicine, Maywood, Illinois, USA.
- Department of Radiation Oncology, Cardinal Bernard Cancer Center, Maywood, Illinois, USA.
- Department of Radiation Oncology, University of Rochester Medical Center, Rochester, New York, USA.
ORCID-Profile: Alexander Podgorsak, Hyejoo Kang
Zitieren
APA
Tsui, T., Podgorsak, A., Roeske, J. C., Small, W., Refaat, T., & Kang, H. (2024). Geometric and dosimetric evaluation for breast and regional nodal auto‐segmentation structures. Journal of Applied Clinical Medical Physics, 25(10). https://doi.org/10.1002/acm2.14461
Vancouver
Tsui T, Podgorsak A, Roeske JC, Small W, Refaat T, Kang H. Geometric and dosimetric evaluation for breast and regional nodal auto‐segmentation structures. Journal of Applied Clinical Medical Physics. 2024;25(10). doi:10.1002/acm2.14461
BibTeX
@article{Tsui2024,
author = {Tsui, Tiffany and Podgorsak, Alexander and Roeske, John C and Small, William and Refaat, Tamer and Kang, Hyejoo},
title = {Geometric and dosimetric evaluation for breast and regional nodal auto‐segmentation structures},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {10},
doi = {10.1002/acm2.14461}
}
EUVIMED