An adult and pediatric size‐based contrast administration reduction phantom study for single and dual‐energy CT through preservation of contrast‐to‐noise ratio

Jia Wang, Xinhui Duan iD, Usman Mahmood, Sarah Eva McKenney, Samuel Loren Brady

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14340 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Jia Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Environmental Health & Safety, Stanford University, Stanford, California, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Contrast Media, Tomography, X-Ray Computed, COVID-19. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Global shortages of iodinated contrast media (ICM) during COVID‐19 pandemic forced the imaging community to use ICM more strategically in CT exams.

Purpose

The purpose of this work is to provide a quantitative framework for preserving iodine CNR while reducing ICM dosage by either lowering kV in single‐energy CT (SECT) or using lower energy virtual monochromatic images (VMI) from dual‐energy CT (DECT) in a phantom study. Materials and Methods In SECT study, phantoms with effective diameters of 9.7, 15.9, 21.1, and 28.5 cm were scanned on SECT scanners of two different manufacturers at a range of tube voltages. Statistical based iterative reconstruction and deep learning reconstruction were used. In DECT study, phantoms with effective diameters of 20, 29.5, 34.6, and 39.7 cm were scanned on DECT scanners from three different manufacturers. VMIs were created from 40 to 140 keV. ICM reduction by lowering kV levels for SECT or switching from SECT to DECT was calculated based on the linear relationship between iodine CNR and its concentration under different scanning conditions.

Results

On SECT scanner A, while matching CNR at 120 kV, ICM reductions of 21%, 58%, and 72% were achieved at 100, 80, and 70 kV, respectively. On SECT scanner B, 27% and 80% ICM reduction was obtained at 80 and 100 kV. On the Fast‐kV switch DECT, with CNR matched at 120 kV, ICM reductions were 35%, 30%, 23%, and 15% with VMIs at 40, 50, 60, and 68 keV, respectively. On the dual‐source DECT, ICM reductions were 52%, 48%, 42%, 33%, and 22% with VMIs at 40, 50, 60, 70, and 80 keV. On the dual‐layer DECT, ICM reductions were 74%, 62%, 45%, and 22% with VMIs at 40, 50, 60, and 70 keV.

Conclusions

Our work provided a quantitative baseline for other institutions to further optimize their scanning protocols to reduce the use of ICM.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingContrast MediaTomography, X-Ray ComputedCOVID-19HumansSARS-CoV-2AdultChildSignal-To-Noise RatioRadiation DosageImage Processing, Computer-AssistedRadiography, Dual-Energy Scanned Projection

Stichwörter der Autor:innen

dual‐energy CTiodinated contrast mediaiterative reconstructionvirtual monochromatic images

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Environmental Health & Safety, Stanford University, Stanford, California, USA.
  • Department of Radiology, UT Southwestern Medical Center, Dallas, Texas, USA.
  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, USA.
  • Department of Radiology, University of California, Davis Medical Center, Sacramento, California, USA.
  • Department of Radiology, Cincinnati Children's Hospital Medical Center, University of Cincinnati, Cincinnati, Ohio, USA.

ORCID-Profile: Xinhui Duan

Förderung

  • NCI NIH HHS

Zitieren

APA

Wang, J., Duan, X., Mahmood, U., McKenney, S. E., & Brady, S. L. (2024). An adult and pediatric size‐based contrast administration reduction phantom study for single and dual‐energy CT through preservation of contrast‐to‐noise ratio. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14340

Vancouver

Wang J, Duan X, Mahmood U, McKenney SE, Brady SL. An adult and pediatric size‐based contrast administration reduction phantom study for single and dual‐energy CT through preservation of contrast‐to‐noise ratio. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14340

BibTeX
@article{Wang2024,
  author = {Wang, Jia and Duan, Xinhui and Mahmood, Usman and McKenney, Sarah Eva and Brady, Samuel Loren},
  title = {An adult and pediatric size‐based contrast administration reduction phantom study for single and dual‐energy CT through preservation of contrast‐to‐noise ratio},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14340}
}
Kostenlos starten →