Intraoperative radiation therapy with 50 kV x‐rays: A multi‐institutional review
Journal of Applied Clinical Medical Physics · 2024-03 · 25 (3)
DOI: 10.1002/acm2.14272 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 26.01.2024. Verfasst von Anil Sethi und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Stritch School of Medicine, Loyola University Chicago, Maywood, Illinois, USA. Hauptthemen laut MEDLINE: Radiometry, In Vivo Dosimetry. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 72 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This report covers clinical implementation of a low kV intraoperative radiation therapy (IORT) program with the INTRABEAM® System (Carl Zeiss Meditec AG, Jena, Germany). Based on collective user experience from eight institutions, we discuss best methods of INTRABEAM quality assurance (QA) tests, commissioning measurements, clinical workflow, treatment planning, and potential avenues for research. The guide provides pertinent background information and clinical justification for IORT. It describes the INTRABEAM system and commissioning measurements along with a TG100 risk management analysis to ensure safety and accuracy of the IORT program. Following safety checks, dosimetry measurements are performed for verification of field flatness and symmetry, x‐ray output, and depth dose. Also discussed are dose linearity checks, beam isotropy, ion chamber measurements, calibration protocols, and in‐vivo dosimetry with optically stimulated luminescence dosimeters OSLDs, and radiochromic film. Emphasis is placed on the importance of routine QA procedures (daily, monthly, and annual) performed at regular intervals for a successful IORT program. For safe and accurate dose delivery, tests of important components of IORT clinical workflow are emphasized, such as, dose prescription, pre‐treatment QA, treatment setup, safety checks, radiation surveys, and independent checks of delivered dose. Challenges associated with in‐vivo dose measurements are discussed, along with special treatment procedures and shielding requirements. The importance of treatment planning in IORT is reviewed with reference to a Monte Carlo‐based commercial treatment planning system highlighting its main features and limitations. The report concludes with suggested topics for research including CT‐based image‐guided treatment planning and improved prescription dose accuracy. We hope that this multi‐institutional report will serve as a guidance document on the clinical implementation and use of INTRABEAM IORT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Stritch School of Medicine, Loyola University Chicago, Maywood, Illinois, USA.
- Department of Radiation Oncology, Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, New York, USA.
- Department of Radiation Oncology, Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California, USA.
- Department of Medical Physics, McGill University Health Center, Montreal, Quebec, Canada.
- Department of Radiation Oncology, Weill Cornell Medical Center, New York-Presbyterian Hospital, New York, New York, USA.
ORCID-Profile: Anil Sethi, Wolfgang Axel Tomé, Xiaofeng Yang, Hualin Zhang
Zitieren
APA
Sethi, A., Gros, S., Brodin, P., Ghavidel, B., Chai, X., Popovic, M., Tomé, W. A., Trichter, S., Yang, X., Zhang, H., & Uhl, V. (2024). Intraoperative radiation therapy with 50 kV x‐rays: A multi‐institutional review. Journal of Applied Clinical Medical Physics, 25(3). https://doi.org/10.1002/acm2.14272
Vancouver
Sethi A, Gros S, Brodin P, Ghavidel B, Chai X, Popovic M, et al. Intraoperative radiation therapy with 50 kV x‐rays: A multi‐institutional review. Journal of Applied Clinical Medical Physics. 2024;25(3). doi:10.1002/acm2.14272
BibTeX
@article{Sethi2024,
author = {Sethi, Anil and Gros, Sebastien and Brodin, Patrik and Ghavidel, Beth and Chai, Xuedong and Popovic, Marija and Tomé, Wolfgang Axel and Trichter, Samuel and Yang, Xiaofeng and Zhang, Hualin and Uhl, Valery},
title = {Intraoperative radiation therapy with 50 kV x‐rays: A multi‐institutional review},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {3},
doi = {10.1002/acm2.14272}
}
EUVIMED