Technical specifications of dose management systems: An international atomic energy agency survey
Journal of Applied Clinical Medical Physics · 2024-01 · 25 (1)
DOI: 10.1002/acm2.14219 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 07.12.2023. Verfasst von Ioannis A Tsalafoutas und 7 weiteren Autor:innen. Beteiligte Einrichtung: Hamad Medical Corporation, Occupational Health and Safety, Medical Physics Section, Doha, Qatar. Hauptthemen laut MEDLINE: Nuclear Energy. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Dose management systems (DMS) have been introduced in radiological services to facilitate patient radiation dose management and optimization in medical imaging. The purpose of this study was to gather as much information as possible on the technical characteristics of DMS currently available, regarding features that may be considered essential for simply ensuring regulatory compliance or desirable to fully utilize the potential role of DMS in optimization of many aspects of radiological examinations.
Methods
A technical survey was carried out and all DMS developers currently available (both commercial and open source) were contacted and were asked to participate. An extensive questionnaire was prepared and uploaded in the IAEA International Research Integration System (IRIS) online platform which was used for data collection process. Most of the questions (93%) required a “Yes/No” answer, to facilitate an objective analysis of the survey results. Some free text questions and comments’ slots were also included, to allow participants to give additional information and clarifications where necessary. Depending on the answer, they were considered either as “Yes” or “No.” Results Given the way that the questions were posed, every positive response indicated that a feature was offered. Thus, the percentage of positive responses was used as a measure of adherence. The percentages of positive answers per section (and sub‐section) are presented in graphs and limitations of this type of analysis are discussed in detail.
Conclusions
The results of this survey clearly exhibit that large differences exist between the various DMS developers. Consequently, potential end users of a DMS should carefully determine which of the features available are essential for their needs, prioritize desirable features, but also consider their infrastructure, the level of support required and the budget available before selecting a DMS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Hamad Medical Corporation, Occupational Health and Safety, Medical Physics Section, Doha, Qatar.
- Sengkang General Hospital, Singapore, Singapore.
- Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, Vienna, Austria.
- Division of Human Health, International Atomic Energy Agency, Vienna, Austria.
- Fundación Médica de Río Negro y Neuquén-Leben Salud, Rio Negro, Argentina.
- Servicio de Física Médica, Hospital Clínico San Carlos, Madrid, Spain.
ORCID-Profile: Ioannis A Tsalafoutas
Zitieren
APA
Tsalafoutas, I. A., Arlany, L., Titovich, E., Pynda, Y., Ruggeri, R., Sánchez, R. M., Reiser, I., & Tsapaki, V. (2024). Technical specifications of dose management systems: An international atomic energy agency survey. Journal of Applied Clinical Medical Physics, 25(1). https://doi.org/10.1002/acm2.14219
Vancouver
Tsalafoutas IA, Arlany L, Titovich E, Pynda Y, Ruggeri R, Sánchez RM, et al. Technical specifications of dose management systems: An international atomic energy agency survey. Journal of Applied Clinical Medical Physics. 2024;25(1). doi:10.1002/acm2.14219
BibTeX
@article{Tsalafoutas2024,
author = {Tsalafoutas, Ioannis A and Arlany, Laurentcia and Titovich, Egor and Pynda, Yaroslav and Ruggeri, Ricardo and Sánchez, Roberto Mariano and Reiser, Ingrid and Tsapaki, Virginia},
title = {Technical specifications of dose management systems: An international atomic energy agency survey},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {1},
doi = {10.1002/acm2.14219}
}
EUVIMED