Comparison of setup accuracy of optical surface image versus orthogonal x‐ray images for VMAT of the left breast using deep‐inspiration breath‐hold

Wei Lu iD, Linda X Hong, Nelson Yamada, Sean L Berry, Yulin Song, Wookjin Choi, Laura I Cerviño, Xiaoli Tang, James G Mechalakos, Paul B Romesser, Simon Powell, Guang Li iD

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14117 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Wei Lu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Breast Neoplasms. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To compare the setup accuracy of optical surface image (OSI) versus orthogonal x‐ray images (2DkV) using cone beam computed tomography (CBCT) as ground truth for radiotherapy of left breast cancer in deep‐inspiration breath‐hold (DIBH). Ten left breast DIBH patients treated with volumetric modulated arc therapy (VMAT) were studied retrospectively. OSI, 2DkV, and CBCT were acquired weekly at treatment setup. OSI, 2DkV, and CBCT were registered to planning CT or planning DRR based on a breast surface region of interest (ROI), bony anatomy (chestwall and sternum), and both bony anatomy and breast surface, respectively. These registrations provided couch shifts for each imaging system. The setup errors, or the difference in couch shifts between OSI and CBCT were compared to those between 2DkV and CBCT. A second OSI was acquired during last beam delivery to evaluate intrafraction motion. The median absolute setup errors were (0.21, 0.27, 0.23 cm, 0.6°, 1.3°, 1.0°) for OSI, and (0.26, 0.24, 0.18 cm, 0.9°, 1.0°, 0.6°) for 2DkV in vertical, longitudinal and lateral translations, and in rotation, roll and pitch, respectively. None of the setup errors was significantly different between OSI and 2DkV. For both systems, the systematic and random setup errors were ≤0.6 cm and ≤1.5° in all directions. Nevertheless, larger setup errors were observed in some sessions in both systems. There was no correlation between OSI and CBCT whereas there was modest correlation between 2DkV and CBCT. The intrafraction motion in DIBH detected by OSI was small with median absolute translations

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedBreast NeoplasmsHumansFemaleRetrospective StudiesX-RaysCone-Beam Computed TomographyRadiotherapy Planning, Computer-AssistedBreath Holding

Stichwörter der Autor:innen

2DkVCBCTDIBHOSIleft breast canceroptical surface imageorthogonal x-ray imageradiotherapysetup accuracy

Autor:innen und Institutionen

  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
  • Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, New York, USA.

ORCID-Profile: Wei Lu, Guang Li

Förderung

  • NCI NIH HHS

Zitieren

APA

Lu, W., Hong, L. X., Yamada, N., Berry, S. L., Song, Y., Choi, W., Cerviño, L. I., Tang, X., Mechalakos, J. G., Romesser, P. B., Powell, S., & Li, G. (2023). Comparison of setup accuracy of optical surface image versus orthogonal x‐ray images for VMAT of the left breast using deep‐inspiration breath‐hold. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14117

Vancouver

Lu W, Hong LX, Yamada N, Berry SL, Song Y, Choi W, et al. Comparison of setup accuracy of optical surface image versus orthogonal x‐ray images for VMAT of the left breast using deep‐inspiration breath‐hold. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14117

BibTeX
@article{Lu2023,
  author = {Lu, Wei and Hong, Linda X and Yamada, Nelson and Berry, Sean L and Song, Yulin and Choi, Wookjin and Cerviño, Laura I and Tang, Xiaoli and Mechalakos, James G and Romesser, Paul B and Powell, Simon and Li, Guang},
  title = {Comparison of setup accuracy of optical surface image versus orthogonal x‐ray images for VMAT of the left breast using deep‐inspiration breath‐hold},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14117}
}
Kostenlos starten →