Validation of a Handheld 6-Lead Device for QT Interval Monitoring in Resource-Limited Settings
JAMA Network Open · 07.06.2024 · 7 (6) · e2415576
DOI: 10.1001/jamanetworkopen.2024.15576 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift JAMA Network Open, online erschienen am 07.06.2024. Verfasst von John Z Metcalfe und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pulmonary and Critical Care Medicine, Zuckerberg San Francisco General Hospital and Trauma Center, University of California, San Francisc…. Hauptthemen laut MEDLINE: Electrocardiography. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 65 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAID NIH HHS.
Abstract (Originalsprache)
Importance Rifampin-resistant tuberculosis treatment regimens require electrocardiographic (ECG) monitoring due to the use of multiple QTc-prolonging agents. Formal 12-lead ECG devices represent a significant burden in resource-constrained clinics worldwide and a potential barrier to treatment scale-up in some settings.
Objective
To evaluate the diagnostic accuracy of a handheld 6-lead ECG device within resource-constrained clinics. Design, Setting, and Participants This diagnostic study was performed within a multicenter, pragmatic (broad eligibility criteria with no exclusions for randomized participants), phase 3 rifampin-resistant tuberculosis treatment trial (BEAT Tuberculosis [Building Evidence for Advancing New Treatment for Tuberculosis]) in South Africa. A total of 192 consecutive trial participants were assessed, and 191 were recruited for this substudy between January 21, 2021, and March 27, 2023. A low proportion (3 of 432 [0.7%]) of all screened trial participants were excluded due to a QTc interval greater than 450 milliseconds. Triplicate reference standard 12-lead ECG results were human calibrated with readers blinded to 6-lead ECG results. Main Outcomes and Measures Diagnostic accuracy, repeatability, and feasibility of a 6-lead ECG device.
Results
A total of 191 participants (median age, 36 years [IQR, 28-45 years]; 81 female participants [42.4%]; 91 participants [47.6%] living with HIV) with a median of 4 clinic visits (IQR, 3-4 visits) contributed 2070 and 2015 12-lead and 6-lead ECG assessments, respectively. Across 170 participants attending 489 total clinic visits where valid triplicate QTc measurements were available for both devices, the mean 12-lead QTc measurement was 418 milliseconds (range, 321-519 milliseconds), and the mean 6-lead QTc measurement was 422 milliseconds (range, 288-574 milliseconds; proportion of variation explained, R 2 = 0.4; P < .001). At a QTc interval threshold of 500 milliseconds, the 6-lead ECG device had a negative predictive value of 99.8% (95% CI, 98.8%-99.9%) and a positive predictive value of 16.7% (95% CI, 0.4%-64.1%). The normal expected range of within-individual variability of the 6-lead ECG device was high (±50.2 milliseconds [coefficient of variation, 6.0%]) relative to the 12-lead ECG device (±22.0 milliseconds [coefficient of variation, 2.7%]). The mean (SD) increase in the 12-lead QTc measurement during treatment was 10.1 (25.8) milliseconds, with 0.8% of clinic visits (4 of 489) having a QTc interval of 500 milliseconds or more. Conclusions and Relevance This study suggests that simplified, handheld 6-lead ECG devices are effective triage tests that could reduce the need to perform 12-lead ECG monitoring in resource-constrained settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Division of Pulmonary and Critical Care Medicine, Zuckerberg San Francisco General Hospital and Trauma Center, University of California, San Francisco.
- Department of Clinical Medicine, University of the Witwatersrand, Johannesburg, South Africa.
- D-Prime LLC, McLean, Virginia.
- Clario Inc, Philadelphia, Pennsylvania.
Förderung
- NIAID NIH HHS
Zitieren
APA
Metcalfe, J. Z., Economou, T., Naufal, F., Kucukosmanoglu, M., Kleiman, R., Phillips, P. P. J., & Conradie, F. (2024). Validation of a Handheld 6-Lead Device for QT Interval Monitoring in Resource-Limited Settings. JAMA Network Open, 7(6), e2415576. https://doi.org/10.1001/jamanetworkopen.2024.15576
Vancouver
Metcalfe JZ, Economou T, Naufal F, Kucukosmanoglu M, Kleiman R, Phillips PPJ, et al. Validation of a Handheld 6-Lead Device for QT Interval Monitoring in Resource-Limited Settings. JAMA Network Open. 2024;7(6):e2415576. doi:10.1001/jamanetworkopen.2024.15576
BibTeX
@article{Metcalfe2024,
author = {Metcalfe, John Z and Economou, Tamsin and Naufal, Fahd and Kucukosmanoglu, Murat and Kleiman, Robert and Phillips, Patrick P J and Conradie, Francesca},
title = {Validation of a Handheld 6-Lead Device for QT Interval Monitoring in Resource-Limited Settings},
journal = {JAMA Network Open},
year = {2024},
volume = {7},
number = {6},
pages = {e2415576},
doi = {10.1001/jamanetworkopen.2024.15576}
}
EUVIMED